Z-0124-2013Class I

HeartSine Samaritan(R) 300/300P PAD devices with serial numbers 0400000501 to 0700032917; 08A00035000 to...

HeartSine Technologies Ltd / initiated 2012-09-13 / Terminated / root cause: design

Heartsine Technologies, Limited began a recall of HeartSine Samaritan(R) 300/300P PAD devices with serial numbers 0400000501 to 0700032917; 08A00035000 to 10A0070753; 10C00200000 to 10C00210106 Product Usage... on . FDA classified it as Class I and gives the reason as "Certain Samaritan 300/300P PAD devices may experience one or both of the following conditions that could affect the ability to deliver therapy to a patient in an SCA event if needed: Issue 1 (On/Off Issue): the device...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0124-2013
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
HeartSine Technologies Ltd
Product
HeartSine Samaritan(R) 300/300P PAD devices with serial numbers 0400000501 to 0700032917; 08A00035000 to 10A0070753; 10C00200000 to 10C00210106 Product Usage: Samaritan 300/300P PAD is indicated for use on victims of sudden cardiac arrest who are exhibiting the following signs: Unconsciousness Not breathing Without circulation
Product code
MKJ Automated External Defibrillators (Non-Wearable)
Reason for recall
Certain Samaritan 300/300P PAD devices may experience one or both of the following conditions that could affect the ability to deliver therapy to a patient in an SCA event if needed: Issue 1 (On/Off Issue): the device may turn itself on and off without input from the user. When this occures, the normal sequence of audible instruction prompts that occurs when the device is turned on will be emit
Root cause tag
design
FDA root cause
Device Design
Action
HeartSine sent an Urgent: Medical Device Recall dated September 2012 to all affected customers. The letter identifed the affected product, problem and actions to be taken. A response form was included with the letter to be completed by customers and returned to the recalling firm's representative. In addition, HeartSine is providing a software upgrade to bring all users up to a more recent version of the software. For questions call 1-877-877-0147 or your service representative.The company issued an additional press release on September 13, 2013 in a further attempt to reach out to end customers that had not yet responded to the companys efforts.
Quantity in commerce
81,982 in total
Distribution
Worldwide Distribution and US (nationwide) in the following states: AK, AL, AP, AR, AZ, CA, CO, CT, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, TN, TX, UT, VA, WA, and WI.
Product codes and lots
Serial numbers 0400000501 to 0700032917; 08A00035000 to 10A0070753; 10C00200000 to 10C00210106

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0124-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0124-2013Class ITerminatedVoluntary: Firm initiated