Boston Scientific, Z Flex 270, Steerable Sheath, Sterile EO, Rx only, Distributed by Boston Scientific...
Greatbatch Medical began a recall of Boston Scientific, Z Flex 270, Steerable Sheath, Sterile EO, Rx only, Distributed by Boston Scientific Corporation, One Boston Scientific Place, Natick, MA... on . FDA classified it as Class II and gives the reason as "During the investigation of four returned Z Flex 270 Steerable Sheaths that were used with cardiac cryoablation balloon catheters, the internal PTFE sheath liners were found to display varying degrees of damage that...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2459-2012
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Greatbatch Medical
- Product
- Boston Scientific, Z Flex 270, Steerable Sheath, Sterile EO, Rx only, Distributed by Boston Scientific Corporation, One Boston Scientific Place, Natick, MA 01760-1537, Manufactured by Greathbatch Medical, 2300 Berkshire Lane North, Minneapolis, MN 55441. To facilitate transvenous introduction of diagnostic / therapeutic catheters into the vasculature and into the chambers of the heart.
- Product code
- DRA Catheter, Steerable
- Reason for recall
- During the investigation of four returned Z Flex 270 Steerable Sheaths that were used with cardiac cryoablation balloon catheters, the internal PTFE sheath liners were found to display varying degrees of damage that could pose an embolic risk to a patient.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Greatbatch Medical sent a "SAFETY ALERT-Z-FLEX-270 STEERABLE SHEATH" letter dated September 6, 2012 to the affected customer. The letter identified the product, problem, and actions to be taken by the customer. The letter advised the customer to communicate with their customers to assure the that the recalled sheath model is not used in conjunction with cryoablation catheters. Contact the firm at 763-951-8214 for questions regarding this issue.
- Quantity in commerce
- 787
- Distribution
- Distributed in MA.
- Product codes and lots
- Q2112752, Q2110939, Q2130137, Q2131250, Q2130212, Q2150143, W2162515, W2163769, W2160626, W2175891, W2172722, W2184492, W2196366, W2184490, W2205843
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2459-2012 | Class II | Terminated | Voluntary: Firm initiated |