Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0412-2013Class II

Spine Frontier Inspan compressors Part Number: 11-60004. Inspan compressors provide pressure to drive spikes...

SPINEFRONTIER, Inc. dba LESspine Innovations / initiated 2012-08-28 / Terminated / root cause: design

SpineFrontier, Inc. began a recall of Spine Frontier Inspan compressors Part Number: 11-60004. Inspan compressors provide pressure to drive spikes of Inspan implants into the spinous process bone... on . FDA classified it as Class II and gives the reason as "Inspan Compressors may break at weld causing a loss of compression". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0412-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
SPINEFRONTIER, Inc. dba LESspine Innovations
Product
Spine Frontier Inspan compressors Part Number: 11-60004. Inspan compressors provide pressure to drive spikes of Inspan implants into the spinous process bone, resulting in fixation of the implant.
Product code
KWP Appliance, Fixation, Spinal Interlaminal
Reason for recall
Inspan Compressors may break at weld causing a loss of compression
Root cause tag
design
FDA root cause
Device Design
Action
Spine Frontier issued e-mail notices to customers on May 14, 2012 advising of the potential for breakage. Units were replaced with weld free design.
Quantity in commerce
17 units
Distribution
Worldwide Distribution-USA (nationwide) including PR and the states of FL, MA, MO, OK, and TX, and the country of Mexico.
Product codes and lots
Lot Numbers: SFP02323, SFP02323R, and SFP02323RR

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0412-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0412-2013Class IITerminatedVoluntary: Firm initiated