Z-0025-2013Class II

The Eon Mini Implantable Pulse Generator (IPG) is a 16-channel, rechargeable, multi-program system that...

Abbott Medical / initiated 2012-09-04 / Terminated / root cause: manufacturing process

Advanced Neuromodulation Systems Inc. began a recall of The Eon Mini Implantable Pulse Generator (IPG) is a 16-channel, rechargeable, multi-program system that allows the connection of one or two leads with four... on . FDA classified it as Class II and gives the reason as "Eon Mini IPGs manufactured in April 2012 could potentially exhibit a sudden, brief surge in stimulation that would be felt by the patient as uncomfortable or painful. SJMs routine testing, monitoring, and analysis of...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0025-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
The Eon Mini Implantable Pulse Generator (IPG) is a 16-channel, rechargeable, multi-program system that allows the connection of one or two leads with four, eight, and 16 electrode leads. It is powered by a hermetically sealed battery within a titanium case. It uses microelectronic circuitry to generate constant current electrical stimulation. Stimulation programs can be delivered as either single stimulation or MultiStim programs depending on patients need. Product Usage: Eon Mini system is indicated as an aid in the management of chronic intractable pain of the trunk and/or limbs including unilateral or bilateral pain associated with the following: failed back surgery syndrome, intractable low back and leg pain.
Product code
LGW Stimulator, Spinal-Cord, Totally Implanted For Pain Relief
Reason for recall
Eon Mini IPGs manufactured in April 2012 could potentially exhibit a sudden, brief surge in stimulation that would be felt by the patient as uncomfortable or painful. SJMs routine testing, monitoring, and analysis of manufacturing processes identified three lots of Eon Mini IPGs that the internal battery has the potential to come into contact with the internal microcontroller board.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
ST. Jude Medical sent an important Medical Device Advisory letter dated September 4, 2012 to all affected customers. The letter identified the affected product, problem and recommended actions to be taken. For questions contact your St. Jude Medical Neuromodulation Division Representative or call the Technical Services Department at 1-800-727-7846.
Quantity in commerce
56 units
Distribution
US Nationwide Distribution
Product codes and lots
Eon Mini - 3788 (IPG), Multiple Serial Numbers

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0025-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0025-2013Class IITerminatedVoluntary: Firm initiated