Philips Digital Diagnost with patient carriage Model: 712050 with serial number range from 00 000 00 to 09...
Philips Healthcare Inc. began a recall of Philips Digital Diagnost with patient carriage Model: 712050 with serial number range from 00 000 00 to 09 000 543. Product Usage: Stationary radiographic... on . FDA classified it as Class II and gives the reason as "Hook does not securely hold the footplate in vertical position, causing the footplate to disengage and may fall on patient or operator". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2494-2012
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- Philips Digital Diagnost with patient carriage Model: 712050 with serial number range from 00 000 00 to 09 000 543. Product Usage: Stationary radiographic System
- Product code
- MQB Solid State X-Ray Imager (Flat Panel/Digital Imager)
- Reason for recall
- Hook does not securely hold the footplate in vertical position, causing the footplate to disengage and may fall on patient or operator
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Philips Healthcare sent an Urgent Field Safety Notice letter dated August 15, 2012 to all affected customers. The letter identified the affected product, problem and the actions to be taken to avoid or minimize the occurrence of the issue. Customers were advised to follow the instruction for Use and that Philips plans to replace the old hook with one that has a new design. For questions contact your local Philips representative 1-800-722-9377.
- Quantity in commerce
- 536 units
- Distribution
- Worldwide Distribution - US Nationwide and the countries of :Australia, Austria, Bahrain, Belgium, Canada, Chile, China, Czech Republic, Denmark, Finland, , France, Germany, Ireland, Italy, Japan, Netherlands, New Zealand, Norway, Poland, Qatar, Romania, Russia, Saudi Arabia, Sweden, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan. Thailand, Turkey,United Arab Emirates and United Kingdom.
- Product codes and lots
- Serial number range (patient carriage) from: 00 000 00 to 09 000 543.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2494-2012 | Class II | Terminated | Voluntary: Firm initiated |