Flow Meter, CM Assembly, Model #27025, Serial # 13686, 13690, 13712, 13793. Flowmeter, CM Assembly...
Accutron Inc began a recall of Flow Meter, CM Assembly, Model #27025, Serial # 13686, 13690, 13712, 13793. Flowmeter, CM Assembly, Model#31970, Serial#13692. Product Usage: Conscious... on . FDA classified it as Class I and gives the reason as "Accutron is recalling the Flowmeter (Analgesia Gas Machine) because when it flows Nitrous Gas without Oxygen gas being turned on, it functions in an incorrect manner". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0017-2013
- Classification
- Class I
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Accutron Inc
- Product
- Flow Meter, CM Assembly, Model #27025, Serial # 13686, 13690, 13712, 13793. Flowmeter, CM Assembly, Model#31970, Serial#13692. Product Usage: Conscious Sedation Unit Analgesia Gas Machine.
- Product code
- BSZ Gas-Machine, Anesthesia
- Reason for recall
- Accutron is recalling the Flowmeter (Analgesia Gas Machine) because when it flows Nitrous Gas without Oxygen gas being turned on, it functions in an incorrect manner.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Accutron sent an Urgent Advisory Notice letter dated April 6, 2012 to all their customers who purchased the Ultra PC% Cabinet Mount Flowmeter. The letter informed the customers of the problem identified and actions to be taken. Customers were instructed to discontinue use or to continue use of the device only with extreme care until a replacement can be installed. Extreme care includes the use of the titration method for nitrous oxide administration coupled with close monitoring of both the patient and the actual gas flow through the flowtubes. For question contact Accutron at (800) 531-2221 or by email at [email protected].
- Quantity in commerce
- 220 units total
- Distribution
- US Nationwide Distribution
- Product codes and lots
- Model#27025: Lot # 73971, 74929. Model#31970: Lot#74142.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0017-2013 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-0018-2013 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-0019-2013 | Class I | Terminated | Voluntary: Firm initiated | ||
| Z-0020-2013 | Class I | Terminated | Voluntary: Firm initiated |