Siemens syngo.plaza radiological image processing system
Siemens Medical Solutions USA, Inc began a recall of Siemens syngo.plaza radiological image processing system on . FDA classified it as Class II and gives the reason as "Siemens became aware of an unintended behavior when using syngo.plaza. In syngo.plaza within the online "Short Time Storage" (STS), some studies are marked with an archived icon, however, the entry in the Archive...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0048-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Siemens syngo.plaza radiological image processing system.
- Product code
- LLZ System, Image Processing, Radiological
- Reason for recall
- Siemens became aware of an unintended behavior when using syngo.plaza. In syngo.plaza within the online "Short Time Storage" (STS), some studies are marked with an archived icon, however, the entry in the Archive Database referring to the physical file may be missing.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- The firm initiated this correction on June 6, 2012 by issuing update instructions SY129/12/S to consignees. A script will be distributed to all potential affected customer systems which will prevent data loss by setting the delete protection on all affected studies.
- Quantity in commerce
- 11
- Distribution
- Nationwide Distribution including the states of MO, FL, MA, NY, WI, GA, CA and WA.
- Product codes and lots
- Model number 10592457, serial numbers 100263, 100401, 100177, 100138, 100174, 100406, 100249, 100201, 100198, 100181, 100146.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0048-2013 | Class II | Terminated | Voluntary: Firm initiated |