Z-0022-2013Class II

3M Steri-Vac Gas Sterilizer/Aerator models 5XL and 8XL. 3M Steri-Vac Gas Sterilizer/Aerator is a compact unit...

3M Company / initiated 2012-09-13 / Terminated / root cause: design

3M Company - Health Care Business began a recall of 3M Steri-Vac Gas Sterilizer/Aerator models 5XL and 8XL. 3M Steri-Vac Gas Sterilizer/Aerator is a compact unit designed to sterilize heat- and/or... on . FDA classified it as Class II and gives the reason as "3M determined that the electrical grounding mechanism for the Steri-Vac 5XL and 8XL gas Sterilizer/Aerator did not meet all of the requirements of the standard in that the electrical wiring, which is insulated and...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0022-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
3M Company
Product
3M Steri-Vac Gas Sterilizer/Aerator models 5XL and 8XL. 3M Steri-Vac Gas Sterilizer/Aerator is a compact unit designed to sterilize heat- and/or moisture-sensitive devices. This gas sterilizer/aerator is intended for indoor use only.
Product code
FLF Sterilizer, Ethylene-Oxide Gas
Reason for recall
3M determined that the electrical grounding mechanism for the Steri-Vac 5XL and 8XL gas Sterilizer/Aerator did not meet all of the requirements of the standard in that the electrical wiring, which is insulated and contained within the sealed door should have a stronger connection to the protective electrical grounding mechansim.
Root cause tag
design
FDA root cause
Component design/selection
Action
3m Healthcare Service Center sent a Urgent Medical Device Field Correction letter dated September 19. 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. 3M will contact you to schedule an appointment for a field correction on any unit in the range of serial numbers provided in the table. 3Ms goal is to service the affected units with minimal disruption to your facility. We apologize for any inconvenience this field correction may cause. Please be assured that 3M is committed to the highest level of safety and customer service and we are proud of our longstanding safety record for 3M" Steri-Vac" Gas Sterilizers. For Further Information If you have any questions, please contact the 3M Health Care Service Line at 1-800-688-5888 or your local 3M representative. .
Quantity in commerce
707 units (61 US, 646 OUS)
Distribution
Worldwide Distribution--USA (nationwide) including the states of AL, CO, CT, GA, IN, KY, MA, MI, MN, MO, MS, NJ, OH, PA, SC, TN, VA, WA and WI., and the countries of Argentina, Barbados, Chile, China, Colombia, Costa Rica, Ecuador, Germany, India, Indonesia, Japan, Korea, Kuwait, Lebanon, Malaysia, Mexico, Morocco, Pakistan, Paraguay, Peru, Philippines, Poland, Republic of Guinea, Russia, South Africa, Taiwan, Thailand, Trinidad, UAE, Venezuela, Vietnam and Zimbabwe.
Product codes and lots
Model 5XL 1-door, serial numbers from 721799 - 722166.  Model 5XL 2-door, serial numbers from 820248 - 820285.  Model 8XL 1-door, serial numbers from 350608 - 350887. Model 8XL 2-door, serial numbers from 450073 - 450105

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0022-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0022-2013Class IITerminatedVoluntary: Firm initiated