Columbus 4-IN-1 Femoral Cutting Guide M3 Instrument (NQ083R)
Aesculap, Inc. began a recall of Columbus 4-IN-1 Femoral Cutting Guide M3 Instrument (NQ083R) on . FDA classified it as Class II and gives the reason as "The manufacturer informed the recalling firm of a potential laser marking error in size identification on two Columbus 4-IN-1 Femoral Cutting Guide Instruments. The guides are correctly dimensioned, but the laser...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0082-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Aesculap Inc
- Product
- Columbus 4-IN-1 Femoral Cutting Guide M3 Instrument (NQ083R)
- Product code
- LXH Orthopedic Manual Surgical Instrument
- Reason for recall
- The manufacturer informed the recalling firm of a potential laser marking error in size identification on two Columbus 4-IN-1 Femoral Cutting Guide Instruments. The guides are correctly dimensioned, but the laser markings have been incorrectly etched.
- Root cause tag
- manufacturing process
- FDA root cause
- Process change control
- Action
- Both consignees were notified verbally on May 15, 2012 and asked to inspect their inventory for the recalled items, remove the mislabeled instruments and return them to the recalling firm. Both mislabeled instruments were returned to the firm. Customer questions were directed to (610) 984-9074.
- Quantity in commerce
- 2
- Distribution
- Within the US, product was distributed to PA and TN.
- Product codes and lots
- NQ083R