Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0663-2013Class II

Merit Medical's Custom Procedural Tray or Kits K12T-03162B & K12T-02641F.Custom Procedural Tray or Kits is a...

Merit Medical Systems, Inc. / initiated 2012-09-14 / Terminated

Merit Medical Systems, Inc. began a recall of Merit Medical's Custom Procedural Tray or Kits K12T-03162B & K12T-02641F.Custom Procedural Tray or Kits is a combination of legally marketed medical devices... on . FDA classified it as Class II and gives the reason as "Custom Procedural Trays/Kits contain 0.9% Sodium Chloride Injection which were recalled because one confirmed customer report where four separate particulate issues were identified in four individual flexible...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0663-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Merit Medical Systems, Inc.
Product
Merit Medical's Custom Procedural Tray or Kits K12T-03162B & K12T-02641F.Custom Procedural Tray or Kits is a combination of legally marketed medical devices placed into one container for the convenience of the user and used within the limits of the product's intended uses.
Product code
NGT Saline, Vascular Access Flush
Reason for recall
Custom Procedural Trays/Kits contain 0.9% Sodium Chloride Injection which were recalled because one confirmed customer report where four separate particulate issues were identified in four individual flexible containers. The particular matter may result acutely in local inflammation, phlebitis and/or generalized low-level allergic response to the particulate and/or embolize to other organs in the
FDA root cause
Under Investigation by firm
Action
Merit Medical sent an "URGENT PRODUCT RECALL NOTICE" dated September 14, 2012 to all affected customers. The letter identified the affected product, description of problem and recommended user actions to be taken. Consignees were requested to ensure that all appropriate personnel are notified of the affected Merit trays so that they can take immediate action. They were instructed to quarantine all affected kits, remove the bag of 0.9% Sodium Chloride and return the affected product to Merit for credit or replacement. Consignees were requested to complete and sign the Customer Response Form and immediately fax a copy of the completed Customer Response Form to: 1-804-416-1031 and mail the completed original Customer Response Form using the provided postage-paid, self-addressed envelope to Merit. Consignees were given product return instructions.
Quantity in commerce
422 total kits/trays
Distribution
Distributed in Mississippi and New Jersey.
Product codes and lots
Tray and Kit - K12T-03162B & K12T-02641F/ Lots #T243169, Exp. 2012-12, T276753, Exp. 2013-04, T282465, Exp. 2013-02, T259844, Exp. 2013-02, T265274, Exp. 2013-02, Chloride Injection Lot 05-201-JT

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0663-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0663-2013Class IITerminatedVoluntary: Firm initiated