Male, 16Fr, Straight-tip Intermittent Urinary Catheters M16. The product is distributed individually or in...

Cure Medical, LLC / initiated 2012-09-19 / Terminated / root cause: labeling and UDI

Cure Medical LLC began a recall of Male, 16Fr, Straight-tip Intermittent Urinary Catheters M16. The product is distributed individually or in intermediate boxes containing 30 individual, sterile... on . FDA classified it as Class II and gives the reason as "The firm initiated the product recall because labeling on some of the intermediate boxes of catheters identify the contents to be M16C, a coud-tip catheters while the actual contents are M16Fr male, 16Fr, straight-tip...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0004-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Cure Medical, LLC
Product
Male, 16Fr, Straight-tip Intermittent Urinary Catheters M16. The product is distributed individually or in intermediate boxes containing 30 individual, sterile packaged catheters (PA/PE film and dialyzing paper pouches). There are 30 catheters packaged in sterile pouches that are placed into an intermediate box, 10 of these intermediate boxes packaged into one case. K072539 The Cure Catheter" device, M16, is an intermittent urinary catheter intended to be used by males for the purpose of bladder drainage. It is manufactured with conventional medical grade, latex-free, biocompatible materials. The straight-tip has been designed to eliminate trauma to the urethra and is provided in 16Fr size in an easy-to-open, sterile, single-use package. The coud-tip catheter, M16C, is designed with a curved tip that makes it easier to pass through the curvature of the prostatic urethra in men. It is also used to pass through urethral strictures. This product is prescribed for people who have difficulty or are unable to pass a straight catheter through their urethra and into their bladder.
Product code
EZD Catheter, Straight
Reason for recall
The firm initiated the product recall because labeling on some of the intermediate boxes of catheters identify the contents to be M16C, a coud-tip catheters while the actual contents are M16Fr male, 16Fr, straight-tip catheters with the pouch and the case box labeled as M16Fr male, 16Fr, straight-tip catheters.
Root cause tag
labeling and UDI
FDA root cause
Labeling mix-ups
Action
Cure Medical LLC sent an Urgent Medical Device Correction and Removal Notification letter dated September 19, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. The letter instructed the customers to: a) Check their shelves for M16C with lot number 120425 and quarantine all affected product. b) Verify the contents (pouch label) are M16C and correspond with the intermediate box label. c) If all match, have the dealer sign and return a supplied verification record. d) If any mismatch is found, contact Cure Medical for assignment of a return goods authorization. e) Cure Medical will replace any incorrect product along with processing costs for this situation to the dealer. If affected product was further distributed customers were asked to identify those customers who received them and notify them of the problem. For questions customers were instructed to call 1-800-570-1778.
Quantity in commerce
420cases (= 4,200 boxes)
Distribution
Nationwide Distribution
Product codes and lots
Total affected product: M16, Lot 120425: 420 cases = 4,200 boxes = 126,000 each. 2017.03 is shown on the product labeling as the expiry date for the product, 4 years and 11 months from the date of packaging. The product, M16, is identified with the following UPC Codes for the individual pouches, intermediate boxes and case labels:  ¿ Pouch Label UPC (M16) = *+M405M16 ¿ Box Label UPC (M16) = *+M405M163I* " Case Label UPC (M16) = *+M405M165K*   Explanation of Lot Number: The lot number represents a planned packaging date of a set of work orders for various products, in the format YYMMDD at the contract manufacturers facility.   At the order planning stage, the packaging date is estimated (based on production cycle time and pre-printed box delivery) and assigned to each work order The lot number represents a planned packaging date of a set of work orders for various products, in the format YYMMDD at the contract manufacturers facility.   At the order planning stage, the packaging date is estimated (based on production cycle time and pre-printed box delivery) and assigned to each work order scheduled for packaging on this estimated date.  Therefore, one lot number may be assigned to multiple catalog numbers. The lot number 120425 was assigned to both M16 and M16C products by the scheduling department  Actual packaging of the M16 occurred from 23 May 2012 through 25 May 2012.  Actual packaging of the M16C occurred from 19 May through 21 May 2012.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0004-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0004-2013Class IITerminatedVoluntary: Firm initiated