Z-0805-2013Class II

Aesculap(R) - Miethke Shunt System, FT040T - Control Reservoir Set w/Distal Catheter and FT075P 0 Ventric...

Aesculap Inc / initiated 2012-06-04 / Terminated / root cause: manufacturing process

Aesculap, Inc. began a recall of Aesculap(R) - Miethke Shunt System, FT040T - Control Reservoir Set w/Distal Catheter and FT075P 0 Ventric Cathet, w/Stylet and RT-ANG Guide Central Nervous... on . FDA classified it as Class II and gives the reason as "The deflector may not move freely on the ventricular catheter. The deflector comes pre-positioned at the 5 cm mark on the catheter. In some cases the surgeon will want to adjust the deflector to a different location on...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0805-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Aesculap Inc
Product
Aesculap(R) - Miethke Shunt System, FT040T - Control Reservoir Set w/Distal Catheter and FT075P 0 Ventric Cathet, w/Stylet and RT-ANG Guide Central Nervous System Shunt and Components intended to shunt cerebrospinal fluid (CSF) from the lateral ventricles of the brain into the peritoneum.
Product code
JXG Shunt, Central Nervous System And Components
Reason for recall
The deflector may not move freely on the ventricular catheter. The deflector comes pre-positioned at the 5 cm mark on the catheter. In some cases the surgeon will want to adjust the deflector to a different location on the catheter before placing it in the ventricle. There is a small percentage of cases where aggressively attempting to adjust the deflector has a potential to cause the catheter
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Aesculap sent an Important Recall Notification letter dated June 4, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to return the contact their local sales representative or customer service department at 1-800-282-9000 to return the affected product. Customers were also instructed to complete the Inventory Sheet attached to the letter with the quantity being returned. If customers could not locate the affected product they should provide an explanation as to why the inventory will not be returned. Customers with questions should call 610-984-9251 or 610-984-9291. For questions regarding this recall call 1-800-234-9179, ext 5067.
Quantity in commerce
11
Distribution
Nationwide Distribution including MI, PA and TX.
Product codes and lots
Lots 4503489878, 4503792749, 4503682927, 4503508385, 4503789879, 4503800248 and 4503682935.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0805-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0805-2013Class IITerminatedVoluntary: Firm initiated