Z-0111-2013Class II

EVICEL Fibrin Sealant (Human) Rx Only. Manufacturer: Omrix Biopharmaceuticals Ltd. MDA blood bank, Sheba...

Ethicon, Inc / initiated 2012-07-02 / Terminated / root cause: manufacturing process

Ethicon, Inc. began a recall of EVICEL Fibrin Sealant (Human) Rx Only. Manufacturer: Omrix Biopharmaceuticals Ltd. MDA blood bank, Sheba Hospital, Ramat-Gan, POB 888, Kiryat Ono 55000 Israel... on . FDA classified it as Class II and gives the reason as "Ethicon and Omrix Biopharmaceuticals have received reports of air or gas embolism occuring in association with the use of spray devices employing a pressure regulator to administer EVICEL; these events appear to be...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0111-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Ethicon, Inc
Product
EVICEL Fibrin Sealant (Human) Rx Only. Manufacturer: Omrix Biopharmaceuticals Ltd. MDA blood bank, Sheba Hospital, Ramat-Gan, POB 888, Kiryat Ono 55000 Israel U.S. License No. 1603. Distributed by: Ethicon, Inc. P.O. Box 151, Somerville, NJ 08876-0151 USA. The Pressure Regulator System is positioned on the floor or on an IV pole and accepts medical gas (compressed air, nitrogen, CO2) from the central hospital system or from medical gas cylinders.
Product code
FMF Syringe, Piston
Reason for recall
Ethicon and Omrix Biopharmaceuticals have received reports of air or gas embolism occuring in association with the use of spray devices employing a pressure regulator to administer EVICEL; these events appear to be related to the use of the spray device at higher-than-recommended pressures and/or in close proximity to the tissue surface. The field correction initiated is to provide the recommend
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
Ethicon sent a notification letter "Urgent Product Information EVICEL Fibrin Sealant (Human) All Lots and Product Codes" dated June 27, 2012, to all affected customers. Notifications were sent via UPS on July 2, 2012. The notification informed customers of the identified risks and included instructions for use. The notification included a business reply card consignees were requested to fill out and return. Customer were notified by a notification titled "Urgent Product Information Omrix Pressure Regulator All Product Codes" that was sent by Ethicon on August 1, 2012, as a follow up to the letter dated June 27, 2012. The notification informed consignees of the identified risks and included instructions for use. Customers were notified that a company representative would visit them to confirm receipt of the notification and to ensure the enclosed label has been affixed correctly on all Omrix Pressure Regulators. For further questions please call 1-877-384-4266.
Quantity in commerce
4800 pressure regulators distributed worldwide; 1500 distributed within the US
Distribution
Worldwide Distribution -- USA (nationwide) and the countries of Canada, European Union, Singapore, Israel, Mexico and Brazil.
Product codes and lots
To include all lots and product codes for EVICEL and Omrix pressure regulators.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0111-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0111-2013Class IITerminatedVoluntary: Firm initiated