EVICEL Fibrin Sealant (Human) Rx Only. Manufacturer: Omrix Biopharmaceuticals Ltd. MDA blood bank, Sheba...
Ethicon, Inc. began a recall of EVICEL Fibrin Sealant (Human) Rx Only. Manufacturer: Omrix Biopharmaceuticals Ltd. MDA blood bank, Sheba Hospital, Ramat-Gan, POB 888, Kiryat Ono 55000 Israel... on . FDA classified it as Class II and gives the reason as "Ethicon and Omrix Biopharmaceuticals have received reports of air or gas embolism occuring in association with the use of spray devices employing a pressure regulator to administer EVICEL; these events appear to be...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0111-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Ethicon, Inc
- Product
- EVICEL Fibrin Sealant (Human) Rx Only. Manufacturer: Omrix Biopharmaceuticals Ltd. MDA blood bank, Sheba Hospital, Ramat-Gan, POB 888, Kiryat Ono 55000 Israel U.S. License No. 1603. Distributed by: Ethicon, Inc. P.O. Box 151, Somerville, NJ 08876-0151 USA. The Pressure Regulator System is positioned on the floor or on an IV pole and accepts medical gas (compressed air, nitrogen, CO2) from the central hospital system or from medical gas cylinders.
- Product code
- FMF Syringe, Piston
- Reason for recall
- Ethicon and Omrix Biopharmaceuticals have received reports of air or gas embolism occuring in association with the use of spray devices employing a pressure regulator to administer EVICEL; these events appear to be related to the use of the spray device at higher-than-recommended pressures and/or in close proximity to the tissue surface. The field correction initiated is to provide the recommend
- Root cause tag
- manufacturing process
- FDA root cause
- Employee error
- Action
- Ethicon sent a notification letter "Urgent Product Information EVICEL Fibrin Sealant (Human) All Lots and Product Codes" dated June 27, 2012, to all affected customers. Notifications were sent via UPS on July 2, 2012. The notification informed customers of the identified risks and included instructions for use. The notification included a business reply card consignees were requested to fill out and return. Customer were notified by a notification titled "Urgent Product Information Omrix Pressure Regulator All Product Codes" that was sent by Ethicon on August 1, 2012, as a follow up to the letter dated June 27, 2012. The notification informed consignees of the identified risks and included instructions for use. Customers were notified that a company representative would visit them to confirm receipt of the notification and to ensure the enclosed label has been affixed correctly on all Omrix Pressure Regulators. For further questions please call 1-877-384-4266.
- Quantity in commerce
- 4800 pressure regulators distributed worldwide; 1500 distributed within the US
- Distribution
- Worldwide Distribution -- USA (nationwide) and the countries of Canada, European Union, Singapore, Israel, Mexico and Brazil.
- Product codes and lots
- To include all lots and product codes for EVICEL and Omrix pressure regulators.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0111-2013 | Class II | Terminated | Voluntary: Firm initiated |