Z-0077-2013Class II
Helion and Xenion surgical lights, 501/01 series KLM lights. The surgical lights are intended to locally...
Hill-Rom Inc. / initiated 2012-09-19 / Terminated / root cause: design
Trumpf Medical Systems, Inc. began a recall of Helion and Xenion surgical lights, 501/01 series KLM lights. The surgical lights are intended to locally illuminate an operating or examination area of the... on . FDA classified it as Class II and gives the reason as "Incidents regarding fractures of the front joint of the spring arm have been reported". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0077-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Hill-Rom Inc.
- Product
- Helion and Xenion surgical lights, 501/01 series KLM lights. The surgical lights are intended to locally illuminate an operating or examination area of the patients body with a high intensity light.
- Product code
- FSY Light, Surgical, Ceiling Mounted
- Reason for recall
- Incidents regarding fractures of the front joint of the spring arm have been reported.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Trumpf sent an "URGENT RECALL NOTICE" dated September 19, 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. A Confirmation of Receipt Form was attached for customers to complete and return. Contact the firm at 888-474-9359 for questions regarding this notice.
- Quantity in commerce
- 275 units
- Distribution
- Nationwide Distribution-USA (nationwide) including the states of AR, CA, CO, CT, FL, GA, IL, KS, MA, MI, MO, NC, NJ, NY, OH, SC, TX, UT, VT and WI.
- Product codes and lots
- Ondal AC2000 serial range of the spring arms requiring rework: 0111 010 xxxxxx to 0111 066 xxxxx (010 to 066 = month 01/year 20000 to month 06/year 2006 xxxxxx = 5 digit serial number).
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0077-2013 | Class II | Terminated | Voluntary: Firm initiated |