"***Electric Systems Foot Control with Direction Only***Rx Only***"
The Anspach Effort, Inc. began a recall of "***Electric Systems Foot Control with Direction Only***Rx Only***" Product Usage: Cutting and shaping bone including bones of the spine and cranium on . FDA classified it as Class II and gives the reason as "The Anspach Effort, Inc. in Palm Beach Gardens, FL is recalling the foot control/foot pedal due to the Foot Control Cable lacking the shield connection from the cable to the connector body". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0075-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- The Anspach Effort, Inc.
- Product
- "***Electric Systems Foot Control with Direction Only***Rx Only***" Product Usage: Cutting and shaping bone including bones of the spine and cranium.
- Product code
- HBE Drills, Burrs, Trephines & Accessories (Simple, Powered)
- Reason for recall
- The Anspach Effort, Inc. in Palm Beach Gardens, FL is recalling the foot control/foot pedal due to the Foot Control Cable lacking the shield connection from the cable to the connector body.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Anspach sent an Urgent Product Removal letter dated January 9, 2012 to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to contact their sales representative to facilitate the removal action by providing a replacement for the affected product. For question contact your local sales representative or the Product Support department.
- Quantity in commerce
- 41 units
- Distribution
- Worldwide Distribution - US (nationwide) and the countries of: Italy, Japan, Switzerland, Turkey and Taiwan.
- Product codes and lots
- Serial Numbers: E28305271001, E28305271002, E28305271003, E28305271004, E2830527100, E28305271006, E28305271007, E28305271008, E2830521009, E28305271010, E29305270901, E29305270902, E29305270903, E29305271101, E29305271102, E29305271103, E29305271104, E29305271105, E29305271106, E29305271107, E29.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0074-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0075-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0076-2013 | Class II | Terminated | Voluntary: Firm initiated |