Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0268-2013Class II

The brand name of the device is AlgiNot, a dental impression material. Product Part Numbers AlgiNot Intra Kit...

Kerr Corporation / initiated 2012-02-21 / Terminated / root cause: manufacturing process

Kerr Corporation began a recall of The brand name of the device is AlgiNot, a dental impression material. Product Part Numbers AlgiNot Intra Kit 34683 AlgiNot Volume Refills 34682 Lot Number... on . FDA classified it as Class II and gives the reason as "The firm recalled the product because it may cure harder than expected within the shelf life, which could result in a material that may be more difficult to remove. Use of this product on patients with loose teeth may...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0268-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Kerr Corporation
Product
The brand name of the device is AlgiNot, a dental impression material. Product Part Numbers AlgiNot Intra Kit 34683 AlgiNot Volume Refills 34682 Lot Number 1-1017 AlgiNot is a Class II Medical Device. The device listing number this product is 0136496 and the establishment registration number for Kerr Corporation is 1815757. Intended as an alternative to traditional alginate materials.
Product code
ELW Material, Impression
Reason for recall
The firm recalled the product because it may cure harder than expected within the shelf life, which could result in a material that may be more difficult to remove. Use of this product on patients with loose teeth may result in possible tooth fracture or tooth extraction.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Kerr Corporation sent "URGENT: MEDICAL DEVICE RECALL" letters to all affected customers. The letters identified the product, problem, and actions to be taken by the customers. US, Canadian, and Rest of World consignees were notified via USPS 1st class mail on June 19, 2012, and Australian consignees on July 5, 2012. Consignees were instructed to complete and return the Acknowledgement Form. Notification of the recall expansion was sent, via USPS 1st class mail, on February 21, 2012 to US, Canadian, Australia and European, consignees; however, once further investigation determined that all unexpired lots of the product were affected, the Alginot recall was expanded a second recall letter was sent to clarify that all lots of AlgiNot FS Volume recall and AlgiNot recall expansion are affected by this expanded recall. Letters were mailed to US, Canadian and Rest of World (ROW) customers on June 19, 2012; Australian and European consignees were notified of the expansion on July 5, 2012. All consignees were instructed to complete the Return Form to acknowledge the updated instructions for use and the added cautionary statement. Consignees not responding are to be contacted at least three times via telephone/fax/e-mail as follow up on the recall.
Quantity in commerce
45 kits and 87 refills
Distribution
Worldwide Distribution-USA (nationwide) and the countries of Canada, Europe, and Australia.
Product codes and lots
¿Product Part Numbers Lot Number  AlgiNot Intra Kit 34683 1-1017  AlgiNot Volume Refills 34682 1-1017

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0268-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0268-2013Class IITerminatedVoluntary: Firm initiated