Z-0188-2013Class III

cobas 4800 BRAF V600 Mutation Test for In Vitro Diagnostic Use Roche Molecular Systems, Inc

Roche Molecular Systems Inc / initiated 2012-07-31 / Terminated / root cause: packaging

Roche Molecular Systems, Inc. began a recall of cobas 4800 BRAF V600 Mutation Test for In Vitro Diagnostic Use Roche Molecular Systems, Inc., Product Usage: The primary use of the cobas 4800 BRAF V600... on . FDA classified it as Class III and gives the reason as "It was identified that BRAF controls dispositioned as "Internal Use Only" were incorrectly packaged into cobas 4800 BRAF V600 Mutation Test CE-IVD kit lot R03607 and released for distribution. Two control batches...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-0188-2013
Classification
Class III
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Roche Molecular Systems Inc
Product
cobas 4800 BRAF V600 Mutation Test for In Vitro Diagnostic Use Roche Molecular Systems, Inc., Product Usage: The primary use of the cobas 4800 BRAF V600 Mutation Test is the detection of the BRAF V600 mutations in DNA extracted from formalin-fixed, paraffin-embedded human melanoma tissue. It is intended to be used as an aid in selecting melanoma patients whose tumors carry BRAF V600 mutations for treatment with Zelboraf (vemurafenib).
Product code
OWD Somatic Gene Mutation Detection System
Reason for recall
It was identified that BRAF controls dispositioned as "Internal Use Only" were incorrectly packaged into cobas 4800 BRAF V600 Mutation Test CE-IVD kit lot R03607 and released for distribution. Two control batches manufactured during two process validation studies are involved: PV7711: BRAF Mutant (MUT) Control PV0853: BRAF Wild Type (WT) Control
Root cause tag
packaging
FDA root cause
Mixed-up of materials/components
Action
Roche sent a Safety Board Notice - Product Bulletin 2012-013 on 31-Jul-2012 to affected customers. The notice identifed the affected product, description of situation and actios to be taken. Customers were instructed to discontinue use, quarantine and discard any remaining kits of the affected product in inventory. Customers were advised to contact their affiliate sales office for replacement. The notice states a new cobas 4800 BRAF V600 Mutation Test lot (R07749) is anticipated to be available on 03-Aug-2012. For questions call +01.908.253.7569.
Quantity in commerce
44 kits
Distribution
Worldwide Distribution - Australia, Austria, Belgium, China, Hungary, Turkey and United Kingdom.
Product codes and lots
Material number 05985595190; Lot number R03607

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0188-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0188-2013Class IIITerminatedVoluntary: Firm initiated