Z-0132-2013Class II

Biosense Webster MobiCath Bi-Directional Guiding Sheath. Small Curve catalog no. D140010, Large Curve catalog...

Greatbatch Medical / initiated 2012-09-24 / Terminated / root cause: design

Greatbatch Medical began a recall of Biosense Webster MobiCath Bi-Directional Guiding Sheath. Small Curve catalog no. D140010, Large Curve catalog no. D140011. The MobiCath Bi-Directional Guiding... on . FDA classified it as Class II and gives the reason as "We are notifying you of a potential interaction between therapeutic catheters and MobiCath Bi-Directional Guiding Sheaths. We became aware of these interactions while investigating recent complaints regarding catheters...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0132-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Greatbatch Medical
Product
Biosense Webster MobiCath Bi-Directional Guiding Sheath. Small Curve catalog no. D140010, Large Curve catalog no. D140011. The MobiCath Bi-Directional Guiding Sheath is a flexible tipped percutaneous catheter introducer (or guiding sheath) designed for gaining access to the coronary systems. The device features adjustable tip geometry through use of a rotating handle to deflect the catheter. This catheter is assembled with a hemostasis valve and a sideport infusion line. The MobiCath Bi-Directional Guiding Sheath is intended for introducing various cardiovascular catheters into the heart. Including the left side of the heart through the interatrial septum.
Product code
DYB Introducer, Catheter
Reason for recall
We are notifying you of a potential interaction between therapeutic catheters and MobiCath Bi-Directional Guiding Sheaths. We became aware of these interactions while investigating recent complaints regarding catheters "self-advancing" while being used in conjunction with the MobiCath Sheaths.
Root cause tag
design
FDA root cause
Device Design
Action
The firm, Greatbatch, Ltd., sent an "URGENT CUSTOMER NOTIFICATION Medical Device - Voluntary Field Safety Notice" dated Monday, September 24, 2012 to its customer. The letter described the product, problem and actions to be taken. The customer was instructed to read the "Description of the Problem" and "Recommendations for Clinical Use" sections carefully; complete and return the attached Field Action Acknowledgement Form via mail to: Biosense Webster, Incl, 15715 Arrow Highway, Irwindale, CA 91706; Attn: Recall Coordinator, or Fax to: (909) 839-8514; pass on this notice to anyone in your facility that needs to be informed; maintain awareness of this communication until information has been incorporated into the device labeling, and to avoid the catheter inside the MobiCath Bi-Directional Guiding Sheath from "self-advancing", physicians should reposition the catheter handle when manipulating the catheter in and out of the sheath. Greatbatch, Ltd. is updating the MobiCath Bi-Directional Guiding Sheath Instructions For Use (IFU) accordingly. Supplemental information will be provided with future shipments of these devices, until the revised IFU is available. For questions related to the issue please contact your Biosense Webster representative or call (800) 729-9010, Monday through Friday from 9:00 AM to 6:00 PM EST. For questions related to the Acknowledgment Form and its return, please contact Field Action Coordinator at (909) 839-8468.
Quantity in commerce
9040
Distribution
Nationwide distribution: CA only.
Product codes and lots
Lot no.: W2046680, W2046681, W2072907, W2078460, W2084872, W2098770, W2098771, W2102124, W2102128, W2102133, W2102134, W2108295, W2108296, W2108298, W2108299, W2108300, W2125688, W2125690, W2132584, W2132588, W2132590, W2138965, W2143146, W2143147, W2143149, W2143152, W2145645, W2145677, W2145680, W2145681, W2151287, W2151305, W2151308, W2156746, W2156748, W2156750, W2156751, W2162494, W2162496, W2162497, W2162498, W2162499, W2167985, W2167986, W2167987, W2167988, W2167990, W2172705, W2172706, W2172707, W2172708, W2172710, W2172712, W2177727, W2177728, W2177729, W2177730, W2177733, W2177734, W2184476, W2184483, W2184484, W2184485, W2184487, W2184489, W2190457, W2190461, W2190462, W2190464, W2190470, W2196349, W2196351, W2196352, W2196355, W2196359, W2196361, W2196363, W2211842, W2211844, W2211846, W2211849, W2211850, W2211851, W2211852, W2211853, W2213887, W2213890, W2213892, W2213896, W2213898, W2213901, W2213982, W2221631, W2222017, W2222020, W2222022, W2222024, W2222027, W2222030, W2224408, W2224412, W2224417, W2224421, W2224423, W2224433, W2229959, W2229960, W2229961, W2229963, W2229965, W2229973, W2229974, W2236558, W2236561, W2236588.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0132-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0132-2013Class IITerminatedVoluntary: Firm initiated