6.5mm Cancellous Bone Screw Intended to be used for the fixation of the acetabular shell or fracture repair
Encore Medical, Lp began a recall of 6.5mm Cancellous Bone Screw Intended to be used for the fixation of the acetabular shell or fracture repair on . FDA classified it as Class II and gives the reason as "A product complaint was received where a size 20 screw was labeled as size 30. After receipt, measurement of the screw determined it to be a size 25. Further investigation also determined that there were other screws...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0139-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Encore Medical, LP
- Product
- 6.5mm Cancellous Bone Screw Intended to be used for the fixation of the acetabular shell or fracture repair.
- Product code
- HWC Screw, Fixation, Bone
- Reason for recall
- A product complaint was received where a size 20 screw was labeled as size 30. After receipt, measurement of the screw determined it to be a size 25. Further investigation also determined that there were other screws labeled as size 25, but a size 30 screw was inside of the package. It was thought that all of these screws were found in-house and were contained, but it was determined later that on
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling mix-ups
- Action
- Encore Medical sent a Urgent Field Safety Notice dated September 28, 2012, to all affected customers. This notice needs to be passed on all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. Please contact Customer Service to place a replacement order and receive an RMA number. Return all affected devices using the RMA number. For further questions please call (512) 834-6255
- Quantity in commerce
- 1 unit
- Distribution
- Nationwide Distribution only to CA.
- Product codes and lots
- Model Number 10-55-025, Lot Number 006A1057.