Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0157-2013Class II

Gish 3/8 x 3/8" Straight Connector with Luer fitting, Item Numbers: EC2160B, EC2160S, Lot Numbers...

LivaNova USA, Inc. / initiated 2012-09-14 / Terminated / root cause: manufacturing process

Sorin Group USA, Inc. began a recall of Gish 3/8 x 3/8" Straight Connector with Luer fitting, Item Numbers: EC2160B, EC2160S, Lot Numbers: 1210100010, 1215100026, 1217000138, Exp. Dates: 4/16/17... on . FDA classified it as Class II and gives the reason as "Due to a molding irregularity, certain lots of connectors have the potential to leak fluid or entrain air at the interface of the luer fitting and mating cap. Entrained air could result in patient injury". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0157-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
LivaNova USA, Inc.
Product
Gish 3/8 x 3/8" Straight Connector with Luer fitting, Item Numbers: EC2160B, EC2160S, Lot Numbers: 1210100010, 1215100026, 1217000138, Exp. Dates: 4/16/17, 5/31/15, 7/31/15. Product Usage: The connectors are indicated for use in connecting flexible tubing used in extracorporeal blood or fluid circuits. The connectors are distributed as sterile stand alones and bulk non-sterile.
Product code
DTL Adaptor, Stopcock, Manifold, Fitting, Cardiopulmonary Bypass
Reason for recall
Due to a molding irregularity, certain lots of connectors have the potential to leak fluid or entrain air at the interface of the luer fitting and mating cap. Entrained air could result in patient injury.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Sorin Group USA, Inc. sent a Product Recall letter to all affected customers by certified mail on September 14, 2012. The letter identified the affected products, problem and actions to be taken. Customers were instructed to discontinue use of the affected product, return all unused product to the firm and to complete and return a Response Form via fax or email. For further questions, contact the firm at 1-800-650-2623.
Quantity in commerce
1,640 units
Distribution
US Nationwide Distribution in the states of: AL, AR, CA, CO, CT, FL, IL, MI, MN, MO, MS, NC, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA and WI.
Product codes and lots
Item Numbers: EC2160B, EC2160S, Lot Numbers: 1210100010, 1215100026, 1217000138, Exp. Dates: 4/16/17, 5/31/15, 7/31/15.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0157-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0157-2013Class IITerminatedVoluntary: Firm initiated
Z-0158-2013Class IITerminatedVoluntary: Firm initiated