Gish 3/8 x 3/8" Straight Connector with Luer fitting, Item Numbers: EC2160B, EC2160S, Lot Numbers...
Sorin Group USA, Inc. began a recall of Gish 3/8 x 3/8" Straight Connector with Luer fitting, Item Numbers: EC2160B, EC2160S, Lot Numbers: 1210100010, 1215100026, 1217000138, Exp. Dates: 4/16/17... on . FDA classified it as Class II and gives the reason as "Due to a molding irregularity, certain lots of connectors have the potential to leak fluid or entrain air at the interface of the luer fitting and mating cap. Entrained air could result in patient injury". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0157-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- LivaNova USA, Inc.
- Product
- Gish 3/8 x 3/8" Straight Connector with Luer fitting, Item Numbers: EC2160B, EC2160S, Lot Numbers: 1210100010, 1215100026, 1217000138, Exp. Dates: 4/16/17, 5/31/15, 7/31/15. Product Usage: The connectors are indicated for use in connecting flexible tubing used in extracorporeal blood or fluid circuits. The connectors are distributed as sterile stand alones and bulk non-sterile.
- Product code
- DTL Adaptor, Stopcock, Manifold, Fitting, Cardiopulmonary Bypass
- Reason for recall
- Due to a molding irregularity, certain lots of connectors have the potential to leak fluid or entrain air at the interface of the luer fitting and mating cap. Entrained air could result in patient injury.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Sorin Group USA, Inc. sent a Product Recall letter to all affected customers by certified mail on September 14, 2012. The letter identified the affected products, problem and actions to be taken. Customers were instructed to discontinue use of the affected product, return all unused product to the firm and to complete and return a Response Form via fax or email. For further questions, contact the firm at 1-800-650-2623.
- Quantity in commerce
- 1,640 units
- Distribution
- US Nationwide Distribution in the states of: AL, AR, CA, CO, CT, FL, IL, MI, MN, MO, MS, NC, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA and WI.
- Product codes and lots
- Item Numbers: EC2160B, EC2160S, Lot Numbers: 1210100010, 1215100026, 1217000138, Exp. Dates: 4/16/17, 5/31/15, 7/31/15.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0157-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0158-2013 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.