Z-0195-2013Class II

ABL80 FLEX CO-OX analyzer, model #393-841 (all software versions prior to 1.35). The ABL80 FLEX CO-OX is a...

Sendx Medical, Inc. / initiated 2010-01-06 / Terminated / root cause: software

Sendx Medical Inc began a recall of ABL80 FLEX CO-OX analyzer, model #393-841 (all software versions prior to 1.35). The ABL80 FLEX CO-OX is a portable, automated analyzer that measures pH, blood... on . FDA classified it as Class II and gives the reason as "The recall was initiated because SenDx Medical Inc. became aware of a software issue in the ABL80 FLEX CO-OX analyzer related to the reporting of the hemoglobin fractional parameters". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0195-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Sendx Medical, Inc.
Product
ABL80 FLEX CO-OX analyzer, model #393-841 (all software versions prior to 1.35). The ABL80 FLEX CO-OX is a portable, automated analyzer that measures pH, blood gases, electrolytes, glucose and oximetry in whole blood. The ABL80 FLEX CO-OX system is intended for use by trained technologists, nurses, physicians and therapists. It is intended for use in a laboratory environment, near patient or point of care setting
Product code
CHL Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph
Reason for recall
The recall was initiated because SenDx Medical Inc. became aware of a software issue in the ABL80 FLEX CO-OX analyzer related to the reporting of the hemoglobin fractional parameters.
Root cause tag
software
FDA root cause
Software design
Action
SenDx Medical sent a recall letter to all their foreign customers on January 6, 2010, to informed them of the ABL80 FLEX CO-OX recall via the Radiometer Field Action Notification (FAN) system. The Field Action Notes informed the customers of the problem identified and the action to be taken. Customers were instructed to submit confirmation fax#1 to confirm that the customer letter has been distributed to all ABL80 CO-OX customers by January 31, 2010. Customers were instructed to submit confirmation fax #2 to confirm that all ABL80 CO-OX analyzers have been upgraded to version 1.35 by 6/30/10. In the U.S., the customer was provided with a software upgrade CD along with a letter advising them to the issues associated with the recall. Customers with questions were instructed to contact TechLine at (800) 736-0600 opt 2.
Quantity in commerce
767 units
Distribution
Worldwide Distribution-- US Distribution including the state of Ohio and the countries of Switzerland, Germany, France, Poland, Austria, Italy, Japan, Australia, Denmark, China, Sweden, Norway, Spain, United Kingdom, Netherlands, and Czech Republic.
Product codes and lots
Serial Numbers: all ABL80 FLEX CO-OX analyzers with sofware versions 1.30, 1.31, 1.32, 1.33, and 1.34.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0195-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0195-2013Class IITerminatedVoluntary: Firm initiated