Z-0167-2013Class II

SCP Pump Control Panel

LivaNova USA, Inc. / initiated 2012-09-20 / Terminated

Sorin Group USA, Inc. began a recall of SCP Pump Control Panel, Product Code: 60-02-50, Serial Numbers: 60S09131, 60S09133-60S09135, 60S09140-60S09142, 60S08954. The SCP is a cardiopulmonary bypass... on . FDA classified it as Class II and gives the reason as "Sorin Group USA, Inc. has issued an Urgent Field Safety Notice regarding certain serial numbers of the Stockert Centrifugal Pump (SCP) Pump Control Panel . An out of specification circuit board component within the...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0167-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
LivaNova USA, Inc.
Product
SCP Pump Control Panel, Product Code: 60-02-50, Serial Numbers: 60S09131, 60S09133-60S09135, 60S09140-60S09142, 60S08954. The SCP is a cardiopulmonary bypass speed control device indicated for use exclusively with the COBE Revolution pump head for speed-controlled pumping through cardiopulmonary bypass circuit for typical durations of six hours or less.
Product code
DWA Control, Pump Speed, Cardiopulmonary Bypass
Reason for recall
Sorin Group USA, Inc. has issued an Urgent Field Safety Notice regarding certain serial numbers of the Stockert Centrifugal Pump (SCP) Pump Control Panel . An out of specification circuit board component within the Panel may lead to a malfunction consisting of a blank or flashing display. The flow and pressure display information may also be incorrect after the malfunction.
FDA root cause
Mixed-up of materials/components
Action
The firm, Sorin Group USA, Inc., sent an "URGENT FIELD SAFETY NOTICE" dated September 20, 2012 to all affected customers by certified mail. The notice described the product, problem and actions to be taken. Customers were instructed to discontinue use of and segregate affected product for return and replacement to the firm; to complete and return the attached Customer Response Form via fax to: 303-467-6502 or email to: [email protected]; and if they have transferred the affected products to a third party, to pass this information to the Sorin Group. Further questions may be directed to Sorin Group Customer Service at 1-800-650-2623.
Quantity in commerce
8 units (2 units USA and 6 units OUS)
Distribution
Worldwide distribution: USA (nationwide) including states of : MI and NY; and countries of: Belgium, Canada, Japan, India, Russia, United Arab Emirates.
Product codes and lots
Product Code: 60-02-50, Serial Numbers: 60S09131, 60S09133-60S09135, 60S09140-60S09142, 60S08954.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0167-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0166-2013Class IITerminatedVoluntary: Firm initiated
Z-0167-2013Class IITerminatedVoluntary: Firm initiated