Oncentra External Beam 4.1. Radiation treatment planning software designed to analyze and plan radiation...

Elekta Limited / initiated 2012-08-21 / Terminated / root cause: software

Elekta, Inc. began a recall of Oncentra External Beam 4.1. Radiation treatment planning software designed to analyze and plan radiation treatments in their dimension for the purpose of... on . FDA classified it as Class II and gives the reason as "During the planning process the beam weights can be changed by the customers. When the beam weights are changed and renumbered and this happens during the dose calculation it is possible to save and reopen the plan...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0118-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Elekta Limited
Product
Oncentra External Beam 4.1. Radiation treatment planning software designed to analyze and plan radiation treatments in their dimension for the purpose of treating patients with cancer.
Product code
MUJ System, Planning, Radiation Therapy Treatment
Reason for recall
During the planning process the beam weights can be changed by the customers. When the beam weights are changed and renumbered and this happens during the dose calculation it is possible to save and reopen the plan without any error messages.
Root cause tag
software
FDA root cause
Software design
Action
Nucletron sent a Customer Information Bulletin 555.00248 to all affected customers. The bulletin identified the product, problem, and actions to be taken by the customers. The bulletin contained information on how an incorrect dose association can be prevented by not renumbering beams during recalculation. A Read Confirmation Form was included for customers to complete and return. Contact your local Nucleton Service Representative for questions regarding this notice.
Quantity in commerce
52 units
Distribution
Nationwide Distribution-including the states of AL, CA, DC, KS, LA, MO, NB, NY, NC, ND, OH, OR, PA, RI, SC, TX, and WA.
Product codes and lots
None as the product is software

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0118-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0118-2013Class IITerminatedVoluntary: Firm initiated