Oncentra External Beam 4.1. Radiation treatment planning software designed to analyze and plan radiation...
Elekta, Inc. began a recall of Oncentra External Beam 4.1. Radiation treatment planning software designed to analyze and plan radiation treatments in their dimension for the purpose of... on . FDA classified it as Class II and gives the reason as "During the planning process the beam weights can be changed by the customers. When the beam weights are changed and renumbered and this happens during the dose calculation it is possible to save and reopen the plan...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0118-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Elekta Limited
- Product
- Oncentra External Beam 4.1. Radiation treatment planning software designed to analyze and plan radiation treatments in their dimension for the purpose of treating patients with cancer.
- Product code
- MUJ System, Planning, Radiation Therapy Treatment
- Reason for recall
- During the planning process the beam weights can be changed by the customers. When the beam weights are changed and renumbered and this happens during the dose calculation it is possible to save and reopen the plan without any error messages.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Nucletron sent a Customer Information Bulletin 555.00248 to all affected customers. The bulletin identified the product, problem, and actions to be taken by the customers. The bulletin contained information on how an incorrect dose association can be prevented by not renumbering beams during recalculation. A Read Confirmation Form was included for customers to complete and return. Contact your local Nucleton Service Representative for questions regarding this notice.
- Quantity in commerce
- 52 units
- Distribution
- Nationwide Distribution-including the states of AL, CA, DC, KS, LA, MO, NB, NY, NC, ND, OH, OR, PA, RI, SC, TX, and WA.
- Product codes and lots
- None as the product is software
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0118-2013 | Class II | Terminated | Voluntary: Firm initiated |