Synthes External Fixation System Tube to Tube Clamp, Part Number 390.007. Intended for use to provide...
Synthes USA HQ, Inc. began a recall of Synthes External Fixation System Tube to Tube Clamp, Part Number 390.007. Intended for use to provide treatment for long bone and pelvic fractures that require... on . FDA classified it as Class II and gives the reason as "The insert provided with the product for the specific lot number indicates that this product is MR safe. This product is MR Conditional and is etched as MR Conditional. The insert is incorrect". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0277-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Synthes GmbH
- Product
- Synthes External Fixation System Tube to Tube Clamp, Part Number 390.007. Intended for use to provide treatment for long bone and pelvic fractures that require external fixation. Specifically, the components can be used for: Stabilization of soft tissues and fractures; Polytrauma/multiple orthopedic trauma; Vertically stable pelvic fractures, or as treatment adjunct for vertically unstable pelvic fractures; Arthrodeses and osteotomies with soft tissue problems - failures of total joints; Neutralization of fractures stabilized with limited internal fixation; Non-unions/septic non-unions; Intra-operative reductions/stabilization tool to assist with indirect reduction; and Unilateral rectilinear bone segment transport or leg lengthening.
- Product code
- KTT Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component
- Reason for recall
- The insert provided with the product for the specific lot number indicates that this product is MR safe. This product is MR Conditional and is etched as MR Conditional. The insert is incorrect.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling mix-ups
- Action
- Synthes sent an "URGENT NOTICE: MEDICAL DEVICE RECALL" letter dated September 28, 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Customers were asked to complete and return a Verification Form via fax to 610-251-9005. Contact the firm at 610-719-5450 for questions regarding this notice.
- Quantity in commerce
- 8
- Distribution
- Nationwide Distribution-including the states of CA, MN, NC, and TN.
- Product codes and lots
- Part number 390.007, Lot number 6800599
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0277-2013 | Class II | Terminated | Voluntary: Firm initiated |