Z-0469-2013Class II

Stryker Orthopaedics FEMORAL OFFSET BUSHING TRITHION REVISION INSTRUMENTS REF 6543-1-600;NON-STERILE...

Howmedica Osteonics Corp / initiated 2012-07-27 / Terminated / root cause: labeling and UDI

Stryker Howmedica Osteonics Corp. began a recall of Stryker Orthopaedics FEMORAL OFFSET BUSHING TRITHION REVISION INSTRUMENTS REF 6543-1-600;NON-STERILE Howmedica Osteonics Corp. 325 Corporate Drive Mahwah, NJ... on . FDA classified it as Class II and gives the reason as "Offset Bushings associated with the lots identified were mismarked: the rotational reference numbers on the Tibial Offset Bushing were marked counterclockwise instead of clockwise; the Femoral Offset Bushing did not...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0469-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Howmedica Osteonics Corp
Product
Stryker Orthopaedics FEMORAL OFFSET BUSHING TRITHION REVISION INSTRUMENTS REF 6543-1-600;NON-STERILE Howmedica Osteonics Corp. 325 Corporate Drive Mahwah, NJ 07430 USA A Subsidiary of Stryker Corp. Stryker France ZAC Satolas Green Pusignan Av de Satolas Green 69881 Meyzieu Cedex France The Offset Bushings are parts of the Triathlon TS system. They are used during trialing to assess the appropriate magnitude and orientation of offset between the Implant and the intramedullary Stem. The feedback from the Offset Bushing allows for the selection of the corresponding Offset Adaptor, which allows for optimal bone coverage, compensating for differences between the Implant and bone canal axes. Made in USA
Product code
LXH Orthopedic Manual Surgical Instrument
Reason for recall
Offset Bushings associated with the lots identified were mismarked: the rotational reference numbers on the Tibial Offset Bushing were marked counterclockwise instead of clockwise; the Femoral Offset Bushing did not include rotational markings.
Root cause tag
labeling and UDI
FDA root cause
Error in labeling
Action
Stryker Orthopaedics sent an "URGENT PRODUCT RECALL" letter dated July 27, 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. A Product Recall Acknowledgement Form was attached for customers to complete and return via fax to 201-831-6069. Contact the firm at 201-972-2100 for assistance with this recall.
Quantity in commerce
7 units
Distribution
Worldwide Distribution.
Product codes and lots
Class I, 510(k) exempt  Catalog Number: 6543-1-600  Lot Number; NE1WR1

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0469-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0469-2013Class IITerminatedVoluntary: Firm initiated
Z-0470-2013Class IITerminatedVoluntary: Firm initiated