Siemens Dimension Vista(R) Alkaline Phosphatase (ALP) Calibrator The ALP CAL is an in vitro diagnostic...
Siemens Healthcare Diagnostics Inc. / initiated 2012-10-10 / Terminated / root cause: manufacturing process
Siemens Healthcare Diagnostics, Inc. began a recall of Siemens Dimension Vista(R) Alkaline Phosphatase (ALP) Calibrator The ALP CAL is an in vitro diagnostic product for the calibration of the Alkaline Phosphatase... on . FDA classified it as Class III and gives the reason as "Correlation slope bias and accuracy shift". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0988-2013
- Classification
- Class III
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- Siemens Dimension Vista(R) Alkaline Phosphatase (ALP) Calibrator The ALP CAL is an in vitro diagnostic product for the calibration of the Alkaline Phosphatase method on the Dimension Vista" System.
- Product code
- JIT Calibrator, Secondary
- Reason for recall
- Correlation slope bias and accuracy shift
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Siemens sent a Urgent Medical Device Correction letter dated October 16, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were instructed to calibrate with newer lots of Dimension Vista(R) ALP Calibrator as they become available, beginning with lot 2GD011. Customers were also asked to discontinue use and discard any remaining inventory of lot 2AD052. Customers further instructed to complete the attached response form and fax to (302) 631-8467, to indicate they had received the information and to facilitate their no charge replacement. Further questions please call (302) 631-6311.
- Quantity in commerce
- 1985 cartons
- Distribution
- Nationwide distribution: USA including Puerto Rico and the states of AL, OH, MO, GA, IL, CT, IA, SC, VA, TX, FL, NY, NC, CA, NM, AZ, CA, MI, NJ, PR, MS, NE, WV, PA, MT, MA, MD, KS, NH, TN, VT, LA, CO, ID, WA, NH, DE and OR.
- Product codes and lots
- Lot 2AD052 Catolog KC330
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0988-2013 | Class III | Terminated | Voluntary: Firm initiated |