Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0430-2013Class II

0.9% Sodium Chloride Injections, USP, 1000ml, VWR/Baxter

bioMerieux, Inc. / initiated 2012-10-17 / Terminated / root cause: manufacturing process

Biomerieux Inc began a recall of 0.9% Sodium Chloride Injections, USP, 1000ml, VWR/Baxter Product Usage: The concentrated sodium chloride is used to prepare inoculum that is analyzed by the... on . FDA classified it as Class II and gives the reason as "The firm inventoried and distributed 0.9% Sodium Chloride Saline bags as 0.45% Sodium Chloride Saline bags for use with the VITEK 2 system". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0430-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
bioMerieux, Inc.
Product
0.9% Sodium Chloride Injections, USP, 1000ml, VWR/Baxter Product Usage: The concentrated sodium chloride is used to prepare inoculum that is analyzed by the VITEK 2 instrument. The VITEK 2 user manual instructs all users to prepare inoculum with 0.45%-0.50% sodium chloride injections, and the system is designed to use 0.45%-0.50% sodium chloride concentration only.
Reason for recall
The firm inventoried and distributed 0.9% Sodium Chloride Saline bags as 0.45% Sodium Chloride Saline bags for use with the VITEK 2 system.
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
Biomerieux sent an Urgent Product Removal Notice letter dated October 16, 2012 to all affected customers via certified mail. The letter identified the affected product, problem and actions to be taken. Customers were instructed to discontinue use of the affected product and immediately discard any remaining inventory. The letter included instructions on receiving product replacement or credit. For questions or concerns contact your local BioMerieux representative.
Quantity in commerce
199 cases/14-1000mL bags
Distribution
Worldwide Distribution - USA (nationwide) including the states of: AL, AZ, CA, CO, CT, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NY, OH, OK, PA, RI, TN, TX, VA and WI. Distribution was also made to Canada, military facilities and other foriegn countries.
Product codes and lots
Lot number C866715, EXP OCT 13

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0430-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0430-2013Class IITerminatedVoluntary: Firm initiated