Z-0345-2013Class II

UniCel Dxl 800 Access Immunoassay System, Part Numbers: 973100, A71456, and A71457. The UniCel DxI...

Siemens Healthcare Diagnostics Inc. / initiated 2011-08-01 / Terminated / root cause: manufacturing process

Beckman Coulter Inc. began a recall of UniCel Dxl 800 Access Immunoassay System, Part Numbers: 973100, A71456, and A71457. The UniCel DxI Immunoassay System is an in vitro diagnostic device used for... on . FDA classified it as Class II and gives the reason as "The recall was initiated because Beckman Coulter has confirmed that the UniCel Dxl and DxC Access Immunoassay Systems had field modifications to the aspirate and dispense probe assemblies". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0345-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
UniCel Dxl 800 Access Immunoassay System, Part Numbers: 973100, A71456, and A71457. The UniCel DxI Immunoassay System is an in vitro diagnostic device used for the quantitative, semi-quantitative or qualitative determination of various analyte concentrations found in human body fluids. The UniCel DxC series of instruments is a DxI Immunoassay System integrated with a chemistry analyzer.
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
The recall was initiated because Beckman Coulter has confirmed that the UniCel Dxl and DxC Access Immunoassay Systems had field modifications to the aspirate and dispense probe assemblies.
Root cause tag
manufacturing process
FDA root cause
Process design
Action
Beckman Coulter sent a field modification was released in August 2011 which included: - Installing a new aspirate and dispense tubing routing bracket. - Replacing existing aspirate and dispense probe tubing with modified tubing to help eliminatemisrouting of aspirate and dispense probe tubing. During installation the FSE informed the customer: -That aspirate and dispense tubing are different lengths. - That probes are not interchangeable with other instruments once tubing has been modified. - To route probe tubing by referring to the labels on the tubing. An updated service document, Routine Maintenance Changes, A82294B was provided to every customer. For questions regarding this recall call 714-961-4483.
Quantity in commerce
1,756 units
Distribution
Worldwide Distribution - USA (nationwide) and Internationally to Canada, AE, AU, BG, BW, CH, CN, CZ, DE, ES, FR, GB, HK, HR, HU, IN, IT, JO, JP, LB, MA, MX, NL, PL, QA, RO, RU, SK, TN, TR, ZA, AO, AR, AT, BE, CO, DZ, FI, IE, IL, KR, KW, LI, NA, NO, NZ, PF, PH, PR, SE, SG, SZ, TH, TW, VE, VN, and YT.
Product codes and lots
Serial Numbers: 600000 through 602389

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0345-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0345-2013Class IITerminatedVoluntary: Firm initiated
Z-0346-2013Class IITerminatedVoluntary: Firm initiated
Z-0347-2013Class IITerminatedVoluntary: Firm initiated