UniCel DxC 880i, 860i, 680i, and 660i Integrated Systems, Part Numbers: A64871, A64903, A64935, A59102. The...
Beckman Coulter Inc. began a recall of UniCel DxC 880i, 860i, 680i, and 660i Integrated Systems, Part Numbers: A64871, A64903, A64935, A59102. The UniCel DxI Immunoassay System is an in vitro... on . FDA classified it as Class II and gives the reason as "The recall was initiated because Beckman Coulter has confirmed that the UniCel Dxl and DxC Access Immunoassay Systems had field modifications to the aspirate and dispense probe assemblies". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0347-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- UniCel DxC 880i, 860i, 680i, and 660i Integrated Systems, Part Numbers: A64871, A64903, A64935, A59102. The UniCel DxI Immunoassay System is an in vitro diagnostic device used for the quantitative, semi-quantitative or qualitative determination of various analyte concentrations found in human body fluids. The UniCel DxC series of instruments is a DxI Immunoassay System integrated with a chemistry analyzer.
- Product code
- CEM Electrode, Ion Specific, Potassium
- Reason for recall
- The recall was initiated because Beckman Coulter has confirmed that the UniCel Dxl and DxC Access Immunoassay Systems had field modifications to the aspirate and dispense probe assemblies.
- Root cause tag
- manufacturing process
- FDA root cause
- Process design
- Action
- Beckman Coulter sent a field modification was released in August 2011 which included: - Installing a new aspirate and dispense tubing routing bracket. - Replacing existing aspirate and dispense probe tubing with modified tubing to help eliminatemisrouting of aspirate and dispense probe tubing. During installation the FSE informed the customer: -That aspirate and dispense tubing are different lengths. - That probes are not interchangeable with other instruments once tubing has been modified. - To route probe tubing by referring to the labels on the tubing. An updated service document, Routine Maintenance Changes, A82294B was provided to every customer. For questions regarding this recall call 714-961-4483.
- Distribution
- Worldwide Distribution - USA (nationwide) and Internationally to Canada, AE, AU, BG, BW, CH, CN, CZ, DE, ES, FR, GB, HK, HR, HU, IN, IT, JO, JP, LB, MA, MX, NL, PL, QA, RO, RU, SK, TN, TR, ZA, AO, AR, AT, BE, CO, DZ, FI, IE, IL, KR, KW, LI, NA, NO, NZ, PF, PH, PR, SE, SG, SZ, TH, TW, VE, VN, and YT.
- Product codes and lots
- Serial Numbers: Included in the corresponding Dxl system.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0345-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0346-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0347-2013 | Class II | Terminated | Voluntary: Firm initiated |