The touchscreen display (model 94267) designed for use with the XPREZZON bedside monitor (model 91393) is...
Spacelabs Healthcare, Llc began a recall of The touchscreen display (model 94267) designed for use with the XPREZZON bedside monitor (model 91393) is width-452 mm (17.2 inches) and height- 368 mm (14.5... on . FDA classified it as Class II and gives the reason as "Potential failure of Spacelabs Healthcare Touchscreen Display, Model 94267 where the alarm lights failed to illuminate upon installation". As of , the recall status on record is Terminated, and MDM tags the root cause as supplier.
Recall record
- Recall number
- Z-0196-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- GE Healthcare
- Product
- The touchscreen display (model 94267) designed for use with the XPREZZON bedside monitor (model 91393) is width-452 mm (17.2 inches) and height- 368 mm (14.5 inches). The depth without the pedestal --- 67.7 mm (2.67 inches). The display is an integrated LCD 19-Inch. The Spacelabs Healthcare Xprezzon Bedside Monitor passively displays data generated by Spacelabs parameter modules.
- Product code
- MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
- Reason for recall
- Potential failure of Spacelabs Healthcare Touchscreen Display, Model 94267 where the alarm lights failed to illuminate upon installation.
- Root cause tag
- supplier
- FDA root cause
- Vendor change control
- Action
- Spacelabs Healthcare sent a URGENT - MEDICAL DEVICE CORRECTION letter dated October 19, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers are advised to advise their staff of this potential failure and advise them to ensure the alarm lights are functional. The firm will contact the consignee at the earliest possible date to schedule a convenient time for the hospital to upgrade the affected product at no cost. Customers can call Technical Support at 1-800-522-7025 and select 2 for any questions about this recall.
- Quantity in commerce
- 169 units distributed (98 in US and 71 Internationally)
- Distribution
- Worldwide Distribution--US (nationwide) including states of Alaska, California, Florida, Illinois, Michigan, South Carolina, and Texas., and the countries of Canada, Chile, France, Libya, Panama, Qatar, Saudi Arabia, Switzerland and United Kingdom
- Product codes and lots
- Serial Number for U.S. Cosignees: PRO-12240004, PRO-12240015, PRO-12240017, PRO-12240018, PRO-12240022, PRO-12240026, PRO-12240028, PRO-12240030, PRO-12240031, PRO-12240032, PRO-12240033, PRO-12240034, PRO-12240036, PRO-12240037, PRO-12240038, PRO-12240040, PRO-12240041, PRO-12240058, PRO-12240059, PRO-12240062, PRO-12240063, PRO-12240077, PRO-12240084, PRO-12240090, pro-12260008, PRO-12260016, pro-12260018, PRO-12260023, pro-12260024, PRO-12260032, pro-12260033, pro-12260035, PRO-12260036, PRO-12260037, pro-12260038, pro-12260041, pro-12260044, pro-12260055, pro-12260059, PRO-12260060, PRO-12260062, pro-12260064, pro-12260065, pro-12260068, pro-12260070, pro-12260071, pro-12260073, PRO-12260076, PRO-12260077, PRO-12260080, PRO-12260081, PRO-12260082, PRO-12260083, PRO-12260085, PRO-12260086, PRO-12260087, PRO-12260088, PRO-12260092, PRO-12260094, PRO-12260096, PRO-12260097, pro-12260121, pro-12260125, pro-12260131, pro-12260164, pro-12260167, pro-12260172, PRO-12260175, PRO-12260183, pro-12260185, pro-12280036, pro-12280049, pro-12280062, pro-12280119, pro-12280121, pro-12280126, pro-12280127, pro-12280131, pro-12280133, pro-12280134, pro-12280136, pro-12280137, pro-12280152, pro-12280155, pro-12280156, pro-12280160, pro-12280161, pro-12280162, pro-12280163, pro-12280173, pro-12280174, pro-12280175, pro-12280177, pro-12280182, pro-12280205, pro-12280232, pro-12280272, pro-12280273. Serial number for International Consignees: PRO-12240056, PRO-12260005, pro-12260006, PRO-12260012, PRO-12260013, pro-12260019, PRO-12260020, PRO-12260021, pro-12260022, pro-12260025, pro-12260026, pro-12260027, pro-12260028, PRO-12260029, PRO-12260030, pro-12260031, PRO-12260040, PRO-12260043, PRO-12260045, PRO-12260046, pro-12260047, PRO-12260051, PRO-12260053, PRO-12260057, PRO-12260075, pro-12260079, PRO-12260111, PRO-12260118, PRO-12260122, pro-12260124, pro-12260128, pro-12260140, pro-12260143, PRO-12260146, PRO-12260147, PRO-12260149, PRO-12260151, PRO-12260152, PRO-12260153, PRO-12260155, PRO-12260156, PRO-12260157, pro-12260158, PRO-12260159, PRO-12260160, PRO-12260161, PRO-12260162, PRO-12260163, PRO-12260169, PRO-12260170, pro-12260171, pro-12260174, PRO-12260176, PRO-12260178, pro-12260179, pro-12260180, pro-12260182, pro-12260189, pro-12260190, pro-12260191, PRO-12260192, pro-12260193, pro-12260195, pro-12260199, pro-12280006, pro-12280023, pro-12280132, pro-12280135, pro-12280141, pro-12280206, pro-12280225.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0196-2013 | Class II | Terminated | Voluntary: Firm initiated |