Freedom Cordless LED Light System. Freedom System, Standard Length, Part Number 910863-1, Serial Numbers...

Metrex Research / initiated 2011-11-17 / Terminated / root cause: labeling and UDI

Kerr Corporation began a recall of Freedom Cordless LED Light System. Freedom System, Standard Length, Part Number 910863-1, Serial Numbers 77001000-77003470 and Freedom System, Long Length... on . FDA classified it as Class II and gives the reason as "The firm recalled the Freedom Cordless LED Headlight because some units were assembled with an incorrect component in the control board electronics that may cause some Freedom Cordless LED Headlight units to become warm...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0402-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Metrex Research
Product
Freedom Cordless LED Light System. Freedom System, Standard Length, Part Number 910863-1, Serial Numbers 77001000-77003470 and Freedom System, Long Length, Part Number 910863-2, Serial Numbers 77001000-77003470. Freedom Cordless LED Light System is a Class I Medical Device. The device listing number this product is E115952. The intended use of this device is to illuminate oral structures and operating areas.
Product code
EBA Light, Surgical Headlight
Reason for recall
The firm recalled the Freedom Cordless LED Headlight because some units were assembled with an incorrect component in the control board electronics that may cause some Freedom Cordless LED Headlight units to become warm to the touch; however, no skin damage will occur.
Root cause tag
labeling and UDI
FDA root cause
Labeling mix-ups
Action
Orascoptic sent an "URGENT: MEDICAL DEVICE RECALL" letter dated November 17, 2011. The letter identified the product, problem, and actions to be taken by the customers. An Acknowledgement Form was included for customers to complete and return to the firm. Contact Orascoptic Customer Care at 608-831-2555 for assistance regarding this notice.
Quantity in commerce
1330 units
Distribution
Worldwide Distribution-USA (nationwide) and the countries of Canada, Australia, and Europe.
Product codes and lots
Freedom System, Standard Length, Part Number 910863-1, Serial Numbers 77001000-77003470 and Freedom System, Long Length, Part Number 910863-2, Serial Numbers 77001000-77003470.   Freedom Cordless LED Light System has no expiration date. The affected units were produced and distributed since February 2011.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0402-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0402-2013Class IITerminatedVoluntary: Firm initiated