Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0414-2013Class II

The brand name of the device is Laser Loupes, a dentalloupe with laser eye protection. Laser Loupes have been...

Metrex Research / initiated 2012-04-10 / Terminated

Kerr Corporation began a recall of The brand name of the device is Laser Loupes, a dentalloupe with laser eye protection. Laser Loupes have been assigned product code 79 FSP and are classified... on . FDA classified it as Class II and gives the reason as "The firm initiated the recall for Laser Loupes because the ink used to mark which laser frequency each laser Loupe protects against may wipe off. The loupes work according to specification, but the marking may not...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0414-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Metrex Research
Product
The brand name of the device is Laser Loupes, a dentalloupe with laser eye protection. Laser Loupes have been assigned product code 79 FSP and are classified as a Class I Medical Devices. Laser Loupes are available with four (4) different filters of different colors. Each colored filter is associated with specific laser frequency ranges. The device listing number for this product is E160359. The intended use of this device is to magnify the operatory field and provide laser eye protection.
Product code
FSP Loupe, Diagnostic/Surgical
Reason for recall
The firm initiated the recall for Laser Loupes because the ink used to mark which laser frequency each laser Loupe protects against may wipe off. The loupes work according to specification, but the marking may not permanently adhere to the device.
FDA root cause
Finished device change control
Action
Customers were notified via USPS 1st class mail on April 10, 2012 to US and Canadian, and Rest of World (ROW) consignees, and on April 27, 2012 to European consignees. The notification letter identifies the product, reason for recall, and instructions to customers:"Please review the table below to verify that the laser loupes being used in your office has the correct filter color. ORASCOPTIC KINDLY REQUESTS YOUR COOPERATION IN COMPLETING AND FAXING BACK THE ENClOSED ACKNOWLEDGEMENT FORM IN ORDER TO CONFIRM YOUR RECEIPT OF THIS FIELD CORRECTION. If you are an authorized Orascoptic distributor, please inform your affected customers of this notification within forty-eight (48) hours of receipt.". Customer contact information was also provided:"Please contact Orascoptic Customer Care at 608.831.2555 or toll free in the United States and Canada at 1-800369- 3698 for more information.".
Quantity in commerce
639 units
Distribution
Worldwide Distribution--USA (nationwide) and the countries of Canada and EU.
Product codes and lots
The Laser Loupes are custom made-to-order products.   There are no general model, catalog, code, lot or serial numbers. Laser Loupes have no expiration date. The affected units were produced and distributed since September 2008.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0414-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0414-2013Class IITerminatedVoluntary: Firm initiated