Z-0398-2013Class II

00780303201 MIS 12/14 Slotted Femoral Head Provisional -3.5X32 00780303202 MIS 12/14 Slotted Femoral Head...

Zimmer GmbH / initiated 2010-10-10 / Terminated / root cause: design

Zimmer, Inc. began a recall of 00780303201 MIS 12/14 Slotted Femoral Head Provisional -3.5X32 00780303202 MIS 12/14 Slotted Femoral Head Provisional +0X32 00780303203 MIS 12/14 Slotted... on . FDA classified it as Class II and gives the reason as "Surgical technique instructions for the Versys and Slotted Provisional Femoral Heads have been enhanced to inform users that the head may become difficult to visually locate if it becomes displaced deep or medial to the...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0398-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
00780303201 MIS 12/14 Slotted Femoral Head Provisional -3.5X32 00780303202 MIS 12/14 Slotted Femoral Head Provisional +0X32 00780303203 MIS 12/14 Slotted Femoral Head Provisional +3.5X32 00780303204 MIS 12/14 Slotted Femoral Head Provisional +7X32 00780303205 MIS 12/14 Slotted Femoral Head Provisional +10.5X32 00780303214 MIS 12/14 Slotted Femoral Head Provisional +7X32 The provisional heads replicate final femoral implant geometry and are used to trial range of motion, joint stability, and determine appropriate femoral head size and offset during total hip arthroplasty.
Product code
HWT Template
Reason for recall
Surgical technique instructions for the Versys and Slotted Provisional Femoral Heads have been enhanced to inform users that the head may become difficult to visually locate if it becomes displaced deep or medial to the iliopsoas tendon and provide non-invasive imaging methods that can be used to locate the head. To reduce incidence of sub-optimal fit between provisional femoral heads and the mat
Root cause tag
design
FDA root cause
Device Design
Action
The firm, Zimmer, sent notification by email to consignees beginning October 10, 2012 with a follow-up letter titled "URGENT MEDICAL DEVICE CORRECTION" dated October 16, 2012 to its consignees/customers. The letter described the product, problem and actions to be taken. The customers were instructed to complete a spreadsheet of current users and contact information and immediately provide this information via email to: [email protected], and review and become familiar with the content and recommendations of the attached letter sent to the customers. For questions call 1-877-946-2761.
Quantity in commerce
56,516 all lots, all sizes
Distribution
Worldwide distribution: USA (nationwide) and countries of: Asia, Barbados, Canada, Europe, Mexico, Puerto Rico, San Juan, and South America.
Product codes and lots
All lots distributed from lot 69998600 (2/20/1996) to present

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0398-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0389-2013Class IITerminatedVoluntary: Firm initiated
Z-0390-2013Class IITerminatedVoluntary: Firm initiated
Z-0391-2013Class IITerminatedVoluntary: Firm initiated
Z-0392-2013Class IITerminatedVoluntary: Firm initiated
Z-0393-2013Class IITerminatedVoluntary: Firm initiated
Z-0394-2013Class IITerminatedVoluntary: Firm initiated
Z-0395-2013Class IITerminatedVoluntary: Firm initiated
Z-0396-2013Class IITerminatedVoluntary: Firm initiated
Z-0397-2013Class IITerminatedVoluntary: Firm initiated
Z-0398-2013Class IITerminatedVoluntary: Firm initiated
Z-0399-2013Class IITerminatedVoluntary: Firm initiated
Z-0400-2013Class IITerminatedVoluntary: Firm initiated