00780304001 MIS 12/14 Slotted Femoral Head Provisional -3.5X36 00780304002 MIS 12/14 Slotted Femoral Head...
Zimmer, Inc. began a recall of 00780304001 MIS 12/14 Slotted Femoral Head Provisional -3.5X36 00780304002 MIS 12/14 Slotted Femoral Head Provisional +0X36 00780304003 MIS 12/14 Slotted... on . FDA classified it as Class II and gives the reason as "Surgical technique instructions for the Versys and Slotted Provisional Femoral Heads have been enhanced to inform users that the head may become difficult to visually locate if it becomes displaced deep or medial to the...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0400-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- 00780304001 MIS 12/14 Slotted Femoral Head Provisional -3.5X36 00780304002 MIS 12/14 Slotted Femoral Head Provisional +0X36 00780304003 MIS 12/14 Slotted Femoral Head Provisional +3.5X36 00780304004 MIS 12/14 Slotted Femoral Head Provisional +7X36 00780304005 MIS 12/14 Slotted Femoral Head Provisional +10.5X36 The provisional heads replicate final femoral implant geometry and are used to trial range of motion, joint stability, and determine appropriate femoral head size and offset during total hip arthroplasty.
- Product code
- HWT Template
- Reason for recall
- Surgical technique instructions for the Versys and Slotted Provisional Femoral Heads have been enhanced to inform users that the head may become difficult to visually locate if it becomes displaced deep or medial to the iliopsoas tendon and provide non-invasive imaging methods that can be used to locate the head. To reduce incidence of sub-optimal fit between provisional femoral heads and the mat
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- The firm, Zimmer, sent notification by email to consignees beginning October 10, 2012 with a follow-up letter titled "URGENT MEDICAL DEVICE CORRECTION" dated October 16, 2012 to its consignees/customers. The letter described the product, problem and actions to be taken. The customers were instructed to complete a spreadsheet of current users and contact information and immediately provide this information via email to: [email protected], and review and become familiar with the content and recommendations of the attached letter sent to the customers. For questions call 1-877-946-2761.
- Quantity in commerce
- 56,516 all lots, all sizes
- Distribution
- Worldwide distribution: USA (nationwide) and countries of: Asia, Barbados, Canada, Europe, Mexico, Puerto Rico, San Juan, and South America.
- Product codes and lots
- All lots distributed from lot 69998600 (2/20/1996) to present
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0389-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0390-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0391-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0392-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0393-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0394-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0395-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0396-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0397-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0398-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0399-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0400-2013 | Class II | Terminated | Voluntary: Firm initiated |