Siemens syngo.plaza Radiological Image Processing System Usage: Radiological Image Processing System
Siemens Medical Solutions USA, Inc began a recall of Siemens syngo.plaza Radiological Image Processing System Usage: Radiological Image Processing System on . FDA classified it as Class II and gives the reason as "Firm has become aware of an unintended behavior when using syngo.plaza and issued an Update Instruction. The Update Instruction SY050/12/S provides a software update and addresses the following issues: - Patient Rename...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0465-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Siemens syngo.plaza Radiological Image Processing System Usage: Radiological Image Processing System
- Product code
- LLZ System, Image Processing, Radiological
- Reason for recall
- Firm has become aware of an unintended behavior when using syngo.plaza and issued an Update Instruction. The Update Instruction SY050/12/S provides a software update and addresses the following issues: - Patient Rename messages for the same patient are being continually processed for several days. This fills up the 'pcvdbsrv' log and potentially affects performance of the system. - Data could
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- The firm, Siemens, sent "Update Instruction SY050/12/S" on June 20, 2012 to its consignees/customers. The Update Instructions described the product, problem and actions to be taken. It also provides a software update. If you have any questions call 610-219-4834 or email: [email protected].
- Quantity in commerce
- 15
- Distribution
- Nationwide distribution: USA including states of: CA, FL, GA, MA, MN, MO, NJ, NY, OH, and WI.
- Product codes and lots
- Model number 10 592 457