Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0289-2013Class III

ABX PENTRA Creatinine 120 CP, Model Number Creatinine Part # A11A01868, Lot Numbers: 203273706, 203273707...

Horiba Instruments Incorporated / initiated 2012-09-14 / Terminated / root cause: manufacturing process

Horiba Instruments, Inc dba Horiba Medical began a recall of ABX PENTRA Creatinine 120 CP, Model Number Creatinine Part # A11A01868, Lot Numbers: 203273706, 203273707, 203273808, 203273709. The ABX PENTRA Creatinine 120... on . FDA classified it as Class III and gives the reason as "This recall was initiated because some customers were experiencing an alarm during Creatinine calibration "REAG RANGE HIGH". This alarm caused the calibration to fail and prevented the user from using the Creatinine...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0289-2013
Classification
Class III
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Horiba Instruments Incorporated
Product
ABX PENTRA Creatinine 120 CP, Model Number Creatinine Part # A11A01868, Lot Numbers: 203273706, 203273707, 203273808, 203273709. The ABX PENTRA Creatinine 120 CP is an in vitro diagnostic assay for the quantitative determination of creatinine in human serum, plasma and urine based on a kinetic method using alkaline picrate (Jaffe method). It is composed of a 27 ml monoreagent cassette. Reagent is a chemical solution with additives. ABX PENTRA Creatinine 120 CP reagent, with associated calibrator and controls, is a diagnostic reagent for quantitative in vitro determination of Creatinine in human serum, plasma and urine based on a kinetic method using alkaline picrate (Jaffe method). Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes. This reagent is intended for use on the ABX PENTRA 400 Clinical Chemistry Analyzer.
Product code
CGX Alkaline Picrate, Colorimetry, Creatinine
Reason for recall
This recall was initiated because some customers were experiencing an alarm during Creatinine calibration "REAG RANGE HIGH". This alarm caused the calibration to fail and prevented the user from using the Creatinine assay.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
HORIBA Medical sent customer communication letters dated October 24, 2011, to all affected customers via Fax with the exception of two communications that were emailed. The response or confirmation of receipt was conducted via telephone and also documented on attached Notification List. Customers were given a list of procedures to follow and asked to complete and fax the Acknowledgement of Receipt Form to +33 (0)4 67 14 15 17. For questions regarding this recall call 949-453-0500, ext 208.
Quantity in commerce
8,503 units
Distribution
Nationwide Distribution
Product codes and lots
Model Number Creatinine Part # A11A01868, Lot Numbers: 203273706, 203273707, 203273808, 203273709.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0289-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0289-2013Class IIITerminatedVoluntary: Firm initiated