Z-0429-2013Class III

Minotrol 16, Model Numbers: 2043748, 2043752, 2043756, Lot Number: MX090(Levels: Low, Normal, and High)...

Horiba Instruments Incorporated / initiated 2012-09-14 / Terminated / root cause: manufacturing process

Horiba Instruments, Inc dba Horiba Medical began a recall of Minotrol 16, Model Numbers: 2043748, 2043752, 2043756, Lot Number: MX090(Levels: Low, Normal, and High). Minotrol 16 is a tri-level control intended for in... on . FDA classified it as Class III and gives the reason as "HORIBA Medical Irvine Technical Support received reports from some customers that have experienced high recover values for Hemoglobin when using Minotrol 16 lot #MX090 (Level High). Customer communication was sent...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0429-2013
Classification
Class III
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Horiba Instruments Incorporated
Product
Minotrol 16, Model Numbers: 2043748, 2043752, 2043756, Lot Number: MX090(Levels: Low, Normal, and High). Minotrol 16 is a tri-level control intended for in vitro diagnostic use in monitoring the accuracy and precision on HORIBA Medical hematology blood cell counters using HORIBA Medical reagents. Minotrol 16 is an in vitro diagnostic reagent composed of human erythrocytes, mammalian leukocytes, and platelets in a plasma-like fluid with preservatives. Minotrol 16 is a stable material that provides a means of determining the accuracy and precision of hematology blood cell counters. It is handled in the same manner as a patient specimen. The assay tables are determined on validated instruments using the appropriate reagents
Product code
JPK Mixture, Hematology Quality Control
Reason for recall
HORIBA Medical Irvine Technical Support received reports from some customers that have experienced high recover values for Hemoglobin when using Minotrol 16 lot #MX090 (Level High). Customer communication was sent 10/01/2010.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
HORIBA Medical sent a Product Information Notice letter dated November 5, 2010 via Fax. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to follow the instructions included with the Product Information Notice. For questions customers were instructed to call the Technical Support Hotline at 1-888-903-5001. For questions regarding this recall call 949-453-0500, ext 208.
Quantity in commerce
22,954
Distribution
Nationwide Distribution
Product codes and lots
Model Numbers: 2043748, 2043752, 2043756, Lot Number: MX090(Levels Low, Normal, and High).

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0429-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0429-2013Class IIITerminatedVoluntary: Firm initiated