The Trilogy Acetabular System Shell with Cluster Holes Porous 56 mm O.D. Sterile The Trilogy Acetabular...

Zimmer GmbH / initiated 2012-09-20 / Terminated / root cause: manufacturing process

Zimmer, Inc. began a recall of The Trilogy Acetabular System Shell with Cluster Holes Porous 56 mm O.D. Sterile The Trilogy Acetabular System is indicated for either cemented or uncemented... on . FDA classified it as Class II and gives the reason as "Zimmer Trilogy Acetabular System Shell with Cluster Holes, Porous,56 MM O.D., sterile; is missing polar hole threads. Zimmer received one complaint in which no injury was reported". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0416-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
The Trilogy Acetabular System Shell with Cluster Holes Porous 56 mm O.D. Sterile The Trilogy Acetabular System is indicated for either cemented or uncemented use In skeletally mature Individuals undergoing primary or revision surgery for rehabilitating hips damaged as a result of noninflammatory degenerative joint disease (NIDJD) or its composite diagnoses of osteoarthritis, avascular necrosis, protrusio acetabuli, traumatic arthritis, slipped capital epiphysis, fused Hip, fracture of the pelvis, and diastrophic variant. The shells with screw holes permit the use of titanium alloy screws to provide additional fixation and security, particularly in those cases where acetabular bone stock is deficient.
Product code
LPH Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
Reason for recall
Zimmer Trilogy Acetabular System Shell with Cluster Holes, Porous,56 MM O.D., sterile; is missing polar hole threads. Zimmer received one complaint in which no injury was reported.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Zimmer sent a Urgent Medical Device Recall Notice dated September 20, 2012, to all affected customers. The letter identified the products the problem and the action needed to be taken by the customer. Customers were instructed to locate quarantine and return affected product to Zimmer Product Service Department, 1777 West center St. Warsaw, IN 46580 for product in the USA. Customers were asked to complete the Inventory Return Certification Form and follow the instructions where to return this form. For further questions please call (574) 372-4807.
Quantity in commerce
24 USA; 71 worldwide
Distribution
Worldwide Distribution--USA (nationwide) including the states of AL, CA, CT, KS, MA, MD, MN, MO, NY, SD, TX and WI and the countries of Bulgaria,Canada, Germany, Italy, Netherlands, Russian Fed, Sweden and UK
Product codes and lots
REF 6200-056-22, lot 62010615

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0416-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0416-2013Class IITerminatedVoluntary: Firm initiated