Z-0437-2013Class II

NexGen CompleteKnee Solution Stemmed Nonaugmentable Tibial component The NexGen system is a semiconstrained...

Zimmer GmbH / initiated 2012-09-21 / Terminated / root cause: manufacturing process

Zimmer, Inc. began a recall of NexGen CompleteKnee Solution Stemmed Nonaugmentable Tibial component The NexGen system is a semiconstrained, nonlinked condylar knee prostheses. The device is... on . FDA classified it as Class II and gives the reason as "Zimmer is initiating a lot specific recall of the NexGen Stemmed Nonaugmentable Tibial Component due to the devices being processed through a manufacturing cleaning operation that was operating outside of the validated...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0437-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
NexGen CompleteKnee Solution Stemmed Nonaugmentable Tibial component The NexGen system is a semiconstrained, nonlinked condylar knee prostheses. The device is indicated for patients with severe knee pain and disability due to rheumatoid arthritis; osteoarthritis; traumatic arthritis; polyarthritis; collagen disorders and/or avascular necrosis of the femoral condyle; post-traumatic loss of joint configuration (particularly when there is patellofemoral erosion, dysfunction or prior patellectomy); moderate valgus, varus or flexion deformities; salvage of previously failed surgical attempts or for a knee in which satisfactory stability in flexion cannot be obtained at the time of surgery.
Product code
JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Reason for recall
Zimmer is initiating a lot specific recall of the NexGen Stemmed Nonaugmentable Tibial Component due to the devices being processed through a manufacturing cleaning operation that was operating outside of the validated parameters. As a result, the devices may contain residual particulate from the manufacturing process. Risks include: Allergic reaction due to the foreign material or late onset
Root cause tag
manufacturing process
FDA root cause
Process change control
Action
Zimmer sent an Urgent Medical Device Recall letter dated September 21, 2012, to all affected customers. The letter identifiedthe product, the problem, and the action to be taken by the customer. Customers were instructed to locate and quarantine the affected product. Customers were to return the affected product along with the completed Inventory Return Certificate Form. If product was further distributed the Recall Letter should be forwarded to their customers. Product was to be returned to USA accounts - Zimmer Product Service Department, 1777 West Center St. Warsaw, IN 46580. If product has been implanted, affected surgeons will be notified. For questions regarding this recall call 574-372-4807.
Quantity in commerce
123 worldwide (33 USA)
Distribution
Worldwide Distribution - USA including AZ,CA, FL, IA, IL, IN, MO, NE, NJ, NY, OH, PA, WA. Internationally to Thailand, Canada, Germany, France, UK, Hungary, Netherlands, Russian Fed, and Sweden,
Product codes and lots
Catalog 00-5986-047-02, lots 62049114 and 62045235.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0437-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0437-2013Class IITerminatedVoluntary: Firm initiated