Z-0790-2013Class II

Oxoid Antimicrobial Susceptibility Test Discs, Aztreonam 30 ug, IVD, REF CT0264B, Oxoid Ltd., UK...

Remel Inc / initiated 2012-08-21 / Terminated / root cause: manufacturing process

Remel Inc began a recall of Oxoid Antimicrobial Susceptibility Test Discs, Aztreonam 30 ug, IVD, REF CT0264B, Oxoid Ltd., UK, [email protected] (Both Remel & Oxoid, Ltd are a part of... on . FDA classified it as Class II and gives the reason as "The firm is recalling the product due to some cartridges may contain individual discs that are not impregnated with antibiotic. The firm notified their customers of the recall with a notification letter and requested a...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0790-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Remel Inc
Product
Oxoid Antimicrobial Susceptibility Test Discs, Aztreonam 30 ug, IVD, REF CT0264B, Oxoid Ltd., UK, [email protected] (Both Remel & Oxoid, Ltd are a part of Thermo Fisher Scientific) An antimicrobial susceptibility test disc is a device that consists of antimicrobic-impregnated paper discs used to measure, by a disc-agar diffusion technique or a disc-broth elution technique, the in vitro susceptibility of most clinically important bacterial pathogens to antimicrobial agents (Aztreonam).
Product code
JTN Susceptibility Test Discs, Antimicrobial
Reason for recall
The firm is recalling the product due to some cartridges may contain individual discs that are not impregnated with antibiotic. The firm notified their customers of the recall with a notification letter and requested a confirmation fax back of receipt.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
ThermoFisher sent an Important Medical Device Product Recall Notice dated August 17, 2012, to all affected customers. The letter identifed the product, the problem, and the action to be taken by the customer. Customers were instructed to review results reported using this lot and consider restesting and/or seek approproiate expert advice. Customers were requested to destroy any remaining inventory of the affected lot and contact Customer Services or their local Oxoid supplier who will issue a replacement. Customer were asked to return the attached "Acknowledgement & Receipt Form by FAX to 1-877-428-1922 or 1-913-895-4190. Customers with questions were instructed to call the Technical Support Department at 800-255-6730 (US) or 913-888-0939 (International). For questions regarding this recall call 913-888-0939.
Quantity in commerce
30
Distribution
Nationwide Distribution including WI, MO, MA, CT, AL, IL, WA, CA, RI, MD, NJ, NC and TX.
Product codes and lots
Lot 1100699

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0790-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0790-2013Class IITerminatedVoluntary: Firm initiated