Z-0659-2013Class II

The TM Ardis Interbody System Inserter Stainless Steel, Rubber, Silicone Non Sterile Rx only. The TM Ardis...

Zimmer GmbH / initiated 2012-09-20 / Terminated / root cause: design

Zimmer, Inc. began a recall of The TM Ardis Interbody System Inserter Stainless Steel, Rubber, Silicone Non Sterile Rx only. The TM Ardis Inserter is intended for delivery of the TM Ardis... on . FDA classified it as Class II and gives the reason as "Zimmer Spine is has received complaints of difficulty in turning the implant release knob of the TM Ardis Inserter after final positioning of the TM Ardis interbody spacer implant. There has been two reports where the...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0659-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
The TM Ardis Interbody System Inserter Stainless Steel, Rubber, Silicone Non Sterile Rx only. The TM Ardis Inserter is intended for delivery of the TM Ardis Implant into the cleared disc space. The implant is secured to the inserter via finger-tightening the knob of the inserter shaft allowing the lateral grasping arms to secure the implant. Once the implant is inserted and its position confirmed radiographically, the inserter is removed from the implant by turning the shaft knob counterclockwise to loosen the lateral grasping arms from the implant.
Product code
MAX Intervertebral Fusion Device With Bone Graft, Lumbar
Reason for recall
Zimmer Spine is has received complaints of difficulty in turning the implant release knob of the TM Ardis Inserter after final positioning of the TM Ardis interbody spacer implant. There has been two reports where the inserter instrument was inadvertently removed from patient with the implant still engaged. The correction includes revised directions explaining the technique to remove the implant
Root cause tag
design
FDA root cause
Device Design
Action
Zimmer sent a "MEDICAL DEVICE CORRECTION" letter dated September 26, 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. A Risk Manager Confirmation Form was attached for customers to complete and return. Questions and concerns should be addressed to Zimmer Spine Customer Service at 866-774-6368. When re-design is approved, Zimmer will remove and replace all affected devices.
Quantity in commerce
170
Distribution
USA nationwide; Europe, Sweden
Product codes and lots
Part number: 96-701-10001; Lot numbers: 61946711, 61946712, 61991588, 62000412,

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0659-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0659-2013Class IITerminatedVoluntary: Firm initiated