Z-0473-2013Class II

Cobas 8000 Modular Analyzer Series, Model/Catalog/Part Number: 05641489001. The Cobas 8000 is a fully...

Roche Diagnostics GmbH / initiated 2012-11-01 / Terminated / root cause: software

Roche Diagnostics Operations, Inc. began a recall of Cobas 8000 Modular Analyzer Series, Model/Catalog/Part Number: 05641489001. The Cobas 8000 is a fully automated system for clinical chemistry analysis on . FDA classified it as Class II and gives the reason as "On cobas c 701 and cobas c 702 modules of the Cobas 8000 series analyzer, a mismatch between the reagent probe positioning and the aspirated reagent volumes under certain conditions can lead to insufficient reagent...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0473-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Roche Diagnostics GmbH
Product
Cobas 8000 Modular Analyzer Series, Model/Catalog/Part Number: 05641489001. The Cobas 8000 is a fully automated system for clinical chemistry analysis.
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
On cobas c 701 and cobas c 702 modules of the Cobas 8000 series analyzer, a mismatch between the reagent probe positioning and the aspirated reagent volumes under certain conditions can lead to insufficient reagent pipetting. This may affect patient sample and controls recovery for the following tests: Alkaline Phosphatase (ALP2), Alanine Aminotransferase (ALT), Bicarbonate (CO2-L), Creatinine (C
Root cause tag
software
FDA root cause
Software design
Action
On 11/01/2012 an "Urgent Medical Device Correction" notification was sent via UPS Ground (receipt required) to consignees. The device issue was described and recommended actions were provided. A Roche Field Application Specialist will install the new e-barcode during a scheduled visit. Customers also will have an option to install the new e-barcodes themselves, following the instructions provided in the notification. A Software update will be released when available. Customers may contact Technical Support at 1-800-428-2336 for more information.
Quantity in commerce
78 analyzers
Distribution
Nationwide Distribution, including the states of AL, AZ, CA, FL, GA, IL, IN, KS, KY, MA, MD, MI, MN, NC, NJ, NY, OH, VA, WA, and WI.
Product codes and lots
Catalog numbers affected: Alkaline Phosphatase acc. to IFCC Gen.2 05166888190; Alanine Aminotransferase acc. to IFCC without pyridoxal phosphate (P5P) activation 05850797190; Bicarbonate Liquid 05446376190; Creatinine Jaff¿ Gen.2 05168597190; C-Reactive Protein Gen.3 05172373190; and Urea/BUN 05171873190. Serial Number(s): 0917-10, 1020-04, 1020-05, 1021-03, 1021-04, 1021-07, 1021-08, 1023-06, 1023-09, 1023-10, 1024-03, 1024-04, 1024-05, 1024-09, 1025-02, 1025-05, 1025-07, 1025-09, 1026-01, 1026-02, 1027-09, 1028-01, 1028-07, 1030-09, 1031-05, 1031-09, 1032-09, 1032-10, 1033-04, 1033-05, 1107-06, 1109-02, 1109-03, 1111-07, 1111-08, 1112-02, 1112-04, 1115-06, 1115-07, 1115-08, 1115-10, 1117-04, 1117-07, 1118-02, 1126-02, 1126-03, 1126-04, 1127-01, 1127-02, 1127-06, 1127-07, 1138-03, 1138-04, 1139-04, 1139-05, 1139-06, 1140-02, 1140-03, 1140-04, 1140-05, 1140-10, 1141-08, 1141-09, 1142-02, 1142-03, 1142-04, 1143-01, 1143-02, 1143-03, 1143-04, 1143-05, 1143-06, 1144-05, 1146-04, 1146-05, 1146-06, 1146-07, and 1146-08.