Ventlab Rescue Med Infant Resuscitator, Manufactured by: Ventlab Corporation, 155 Boyce Drive, Mockville, NC...

Salter Labs / initiated 2012-07-11 / Terminated / root cause: design

Ventlab Corporation began a recall of Ventlab Rescue Med Infant Resuscitator, Manufactured by: Ventlab Corporation, 155 Boyce Drive, Mockville, NC 27028. The Disposable Resuscitators with optional... on . FDA classified it as Class I and gives the reason as "Possible volume leakage through the inlet valve during compressions of the resuscitator bag". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0179-2013
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Salter Labs
Product
Ventlab Rescue Med Infant Resuscitator, Manufactured by: Ventlab Corporation, 155 Boyce Drive, Mockville, NC 27028. The Disposable Resuscitators with optional built-in pressure monitor is intended to provide respiratory support in the presence of reversible apnea, commonly associated with cardiopulmonary arrest.
Product code
BTM Ventilator, Emergency, Manual (Resuscitator)
Reason for recall
Possible volume leakage through the inlet valve during compressions of the resuscitator bag.
Root cause tag
design
FDA root cause
Component design/selection
Action
VentLab Corporation sent an Attention: Recall Notification letter dated July 10, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to check their inventory for the affected product, complete and return the enclosed response form by fax to 336-753-5002, indicating whether or not their facility has any of the affected product. Customers can also scan and email the response form to [email protected]. For questions regarding this recall call 336-753-5000.
Quantity in commerce
14,602 total for all units
Distribution
Nationwide Distribution including IL, OH, NY and TN.
Product codes and lots
Product Code: BVM700, Lot #101638.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0179-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0170-2013Class ITerminatedVoluntary: Firm initiated
Z-0171-2013Class ITerminatedVoluntary: Firm initiated
Z-0172-2013Class ITerminatedVoluntary: Firm initiated
Z-0173-2013Class ITerminatedVoluntary: Firm initiated
Z-0174-2013Class ITerminatedVoluntary: Firm initiated
Z-0175-2013Class ITerminatedVoluntary: Firm initiated
Z-0176-2013Class ITerminatedVoluntary: Firm initiated
Z-0177-2013Class ITerminatedVoluntary: Firm initiated
Z-0178-2013Class ITerminatedVoluntary: Firm initiated
Z-0179-2013Class ITerminatedVoluntary: Firm initiated
Z-0180-2013Class ITerminatedVoluntary: Firm initiated
Z-0181-2013Class ITerminatedVoluntary: Firm initiated