GE Healthcare Optima 200amx Mobile general purpose radiographic imaging of the human head and body
GE Healthcare, LLC began a recall of GE Healthcare Optima 200amx Mobile general purpose radiographic imaging of the human head and body on . FDA classified it as Class II and gives the reason as "It was discovered GE Healthcare has become aware of a firmware issue on the mobile radiographic product, Brivo XR285amx, Optima XR200amx, Optima XR22Oamx and Optima XR200amx with Digital Upgrade, related to Technique...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0199-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- GE Healthcare
- Product
- GE Healthcare Optima 200amx Mobile general purpose radiographic imaging of the human head and body.
- Product code
- OXO Image-Intensified Fluoroscopic X-Ray System, Mobile
- Reason for recall
- It was discovered GE Healthcare has become aware of a firmware issue on the mobile radiographic product, Brivo XR285amx, Optima XR200amx, Optima XR22Oamx and Optima XR200amx with Digital Upgrade, related to Technique Accuracy as mandated by the US Code of Federal Regulations (21CFR) and IEC 60601-2-54.
- Root cause tag
- software
- FDA root cause
- Radiation Control for Health and Safety Act
- Action
- GE Healthcare will remedy the defect or bring the product into compliance with each applicable Federal and IEC standard. A Service Representative will update the firmware on the system to address the issue. Field Modification Instructions (FMI) 10870, 10872, and 10873 are describing the rework plan the GEHC Engineers will make to the imaging systems compliant with the applicable performance standard. Further questions please call (262) 513-4122.
- Quantity in commerce
- 364 total units installed in US
- Distribution
- Nationwide Distribution
- Product codes and lots
- part numbers: 5555000-3, 5555000-4
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0197-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0198-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0199-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0200-2013 | Class II | Terminated | Voluntary: Firm initiated |