Medical Surgical Bed, Model 3002SEX Medical Surgical Bed The S3" MedSurg Bed, is intended to support and...

Stryker Medical / initiated 2012-12-18 / Terminated / root cause: manufacturing process

Stryker Medical Division of Stryker Corporation began a recall of Medical Surgical Bed, Model 3002SEX Medical Surgical Bed The S3" MedSurg Bed, is intended to support and transport patients within the Med/Surg and Critical... on . FDA classified it as Class II and gives the reason as "Stryker Medical Model 3002S3EX and Model 3005S3EX beds built between March 15, 2012 and May 23, 2012. The affected beds may have been built with brake rings that do not meet specifications, negatively impacting the...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0628-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Stryker Medical
Product
Medical Surgical Bed, Model 3002SEX Medical Surgical Bed The S3" MedSurg Bed, is intended to support and transport patients within the Med/Surg and Critical Care hospital environments. The bed is typically used in pre-op, post-op and recovery areas of hospital facilities. The intended user for this product is both Health Care Providers (HCPs: nurses, nurses aides, and medical doctors) and human patients. Lockout features may limit patient accessible controls
Product code
FNL Bed, Ac-Powered Adjustable Hospital
Reason for recall
Stryker Medical Model 3002S3EX and Model 3005S3EX beds built between March 15, 2012 and May 23, 2012. The affected beds may have been built with brake rings that do not meet specifications, negatively impacting the holding force of the brakes.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Stryker sent an Urgent Medical Device Correction letterdated December 18, 2012, to all affected customer. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to locate the affected product and ensure that the brakes are holding properly. Consignees were directed to contact Stryker (1-800-327-0770, option 2) to make arrangements to replace all brake rings and continue to routinely inspect brake functionality. Customers were asked to Fax (269-488-8691) the enclosed acknowledgment and Stryker field service will contact customers to schedule service. Consignees were directed to forward a copy of this notice to the new users and advise us of their new location. For questions regarding this recall call 269-488-8691.
Quantity in commerce
9
Distribution
Worldwide Distribution - Nationwide and Canada
Product codes and lots
Model 3002SEX, Serial Number 120315371-120315379

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0628-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0628-2013Class IITerminatedVoluntary: Firm initiated
Z-0629-2013Class IITerminatedVoluntary: Firm initiated