DPM 6/7 Patient Monitor Manufactured in China for : Mindray DS USA, Inc. 800 MacArthur Blvd. Mahwah, NJ 07430...
Mindray DS USA, Inc. d.b.a. Mindray North America began a recall of DPM 6/7 Patient Monitor Manufactured in China for : Mindray DS USA, Inc. 800 MacArthur Blvd. Mahwah, NJ 07430 USA. Intended to be used for monitoring... on . FDA classified it as Class II and gives the reason as "There is an issue with the DPM 6 and DPM 7 monitors where the following features may be unavailable : Full Disclosure, Drug, Hemodynamic Calculations, Renal Calculations, Oxygenation Calculations, and Ventilation...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0513-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Datascope Corp.
- Product
- DPM 6/7 Patient Monitor Manufactured in China for : Mindray DS USA, Inc. 800 MacArthur Blvd. Mahwah, NJ 07430 USA. Intended to be used for monitoring, displaying, reviewing, storing and alarming of multiple physiological parameters including ECG (3-lead or 5-lead or 12-lead selectable). arrhythmia detection, ST Segment analysis, heart rate (HR) , respiration rate (RESP), temperature (TEMP), pulse oxygen saturation (SpO2), pulse rate (PR), non-invasive blood pressure (NIBP), invasive blood pressure (IBP), carbon dioxide (CO2), anesthetic gas (AG),impedance cardiograph (ICG), bispectral index (BIS), cardiac output (C.O.) and respiration mechanics (RM).
- Product code
- MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
- Reason for recall
- There is an issue with the DPM 6 and DPM 7 monitors where the following features may be unavailable : Full Disclosure, Drug, Hemodynamic Calculations, Renal Calculations, Oxygenation Calculations, and Ventilation Calculations.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Mindray sent a "PRODUCT CORRECTIVE ACTION" letter dated August 11, 2010 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Customers should contact their Mindray Service Representative at 1-800-288-2121 for assistance with this issue.
- Quantity in commerce
- 124 units
- Distribution
- Nationwide Distribution-USA (nationwide) including the states of WA, IL, FL, MN, SC, NV, MI, TN, PA, LA, NE, TX, and IA.
- Product codes and lots
- K092449 DPM 6- p/n 6802F-PA00001 DPM 7 - p/n 6800F-PA00001
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0513-2013 | Class II | Terminated | Voluntary: Firm initiated |