Z-0630-2013Class II

DPM Central Station Mindray DS USA, Inc. 800 MacArthur Boulevard, Mahwah, NJ 07430 The Central Monitoring...

Datascope Corp. / initiated 2011-05-12 / Terminated / root cause: software

Mindray DS USA, Inc. d.b.a. Mindray North America began a recall of DPM Central Station Mindray DS USA, Inc. 800 MacArthur Boulevard, Mahwah, NJ 07430 The Central Monitoring Station (CMS) network transfers information between... on . FDA classified it as Class II and gives the reason as "There is a software anomaly with the DPM Central Monitoring System where the trend data of one patient maybe replaced with the trend data of a second patient". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0630-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Datascope Corp.
Product
DPM Central Station Mindray DS USA, Inc. 800 MacArthur Boulevard, Mahwah, NJ 07430 The Central Monitoring Station (CMS) network transfers information between DOM Central Station and other network devices. It also allows information transfer between several CMS. Network connections consist of hardwired network cables and or WLAN connections. The DPM Central Station is used for remote monitor management, storing, printing, reviewing, or processing of information from net worked devices, and it is operated by medical personnel in hospitals or medical institutions.
Product code
MSX System, Network And Communication, Physiological Monitors
Reason for recall
There is a software anomaly with the DPM Central Monitoring System where the trend data of one patient maybe replaced with the trend data of a second patient.
Root cause tag
software
FDA root cause
Software design
Action
Mindray sent a Product Corrective Action letter dated May 12, 2012, to all affected customers via certified mail return receipt requested to two medical facilities. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed that the issue would be corrected through a softweare upgrade of the affected product. A Mindray Service Representative would contact their facility to arrange for the software upgrade at no cost to them. For questions regarding this recall call 1-800-288-2121.
Quantity in commerce
two units
Distribution
Nationwide Distribution including OH and GA
Product codes and lots
510 K K080192 P/N 300BF-PA200002

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0630-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0630-2013Class IITerminatedVoluntary: Firm initiated