Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0298-2013Class II

Product is Full OSSEOTITE Parallel Walled Certain Implant, Model Number IFOA411. The product is sterile...

ZimVie US Corp LLC / initiated 2012-09-28 / Terminated / root cause: manufacturing process

Biomet 3i, LLC began a recall of Product is Full OSSEOTITE Parallel Walled Certain Implant, Model Number IFOA411. The product is sterile. Product provides a means for prosthetic attachment in... on . FDA classified it as Class II and gives the reason as "Biomet 3i recalled Full OSSEOTITE 2 Tapered Certain Implant and Full OSSEOTITE Parallel Walled Certain Implant due to a manufacturing condition where the depth of the implant's internal hex is too shallow. This...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0298-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
ZimVie US Corp LLC
Product
Product is Full OSSEOTITE Parallel Walled Certain Implant, Model Number IFOA411. The product is sterile. Product provides a means for prosthetic attachment in single tooth restorations and in partially or fully edentulous spans with multiple single teeth utilizing delayed or immediate loading or with a terminal or intermediary abutment for fixed or removable bridgework and to retain over dentures.
Product code
DZE Implant, Endosseous, Root-Form
Reason for recall
Biomet 3i recalled Full OSSEOTITE 2 Tapered Certain Implant and Full OSSEOTITE Parallel Walled Certain Implant due to a manufacturing condition where the depth of the implant's internal hex is too shallow. This condition may prevent the driver or abutment from fully engaging in the implant.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Biomet 3i sent an "URGENT: MEDICAL DEVICE RECALL" letter dated September 28, 2012. The letter identified the product, problem, and actions to be taken by the customers. Contact the firm at 1-800-342-5454 for questions regarding this notice.
Quantity in commerce
76 devices
Distribution
Worldwide Distribution-USA (nationwide) including the states of VA, SC, HI, LA, OH, FL, IL, NJ, NC, CA, CT, MI, TN, AK, NH, NY, MA, GA, PA, and KY and the countries of United Kingdom, Spain, Germany, Canada, Japan, Singapore, Taiwan, Israel, Poland, Hong Kong, Ireland, The Netherlands, Austria, and France.
Product codes and lots
Model Number IFOA411, Lot # 2011111333, 2011111333-S1

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0298-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0297-2013Class IITerminatedVoluntary: Firm initiated
Z-0298-2013Class IITerminatedVoluntary: Firm initiated