Trilogy F/M Acetabular Shell with Uni-Hole. The Trilogy Acetabular System is indicated for either cemented or...
Zimmer, Inc. began a recall of Trilogy F/M Acetabular Shell with Uni-Hole. The Trilogy Acetabular System is indicated for either cemented or uncemented use in individuals undergoing surgery... on . FDA classified it as Class II and gives the reason as "Some units of Trilogy shells manufactured since March 2009 may be below the lower range of the pore size specification (as low as 28% vs. 30% porosity specification). On 10/31/2012, the firm expanded the recall action...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0771-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- Trilogy F/M Acetabular Shell with Uni-Hole. The Trilogy Acetabular System is indicated for either cemented or uncemented use in individuals undergoing surgery for rehabilitating hip damage.
- Product code
- LPH Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
- Reason for recall
- Some units of Trilogy shells manufactured since March 2009 may be below the lower range of the pore size specification (as low as 28% vs. 30% porosity specification). On 10/31/2012, the firm expanded the recall action to include lots of Uni-Hole Trilogy Acetabular System shells.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Zimmer sent an "URGENT MEDICAL DEVICE NOTIFICATION" letter dated September 20, 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers.The firm expanded the scope of the recall in January 2013 to include devices that use a secondary mean of fixation. New recall letters will be submitted to distributors and surgeons.
- Quantity in commerce
- 275,952 units
- Distribution
- Worldwide Distribution-USA (nationwide) and the countries of Denmark, Germany, Greece, Hungary, Ireland, Italy, Netherlands, Spain, Sweden, UK, Poland, France, Australia, Chile, Canada, India, Japan, New Zealand, Peru, South Africa, Colombia, Ukraine, and UAE.
- Product codes and lots
- 61282267 61633887 61260645 61237937 61500763 61397941 61584769 61199849 61429445 61412089 00369745
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0769-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0770-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0771-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0772-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0773-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0774-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0775-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0776-2013 | Class II | Terminated | Voluntary: Firm initiated |