Z-0451-2013Class II

Kimberly-Clark Patient Warming System - Model 1000 Control Unit The product is a thermal regulating system...

Kimberly-Clark Corporation / initiated 2010-09-16 / Terminated / root cause: design

Kimberly-Clark Corporation began a recall of Kimberly-Clark Patient Warming System - Model 1000 Control Unit The product is a thermal regulating system, indicated for monitoring and controlling patient... on . FDA classified it as Class II and gives the reason as "Error 85 alarm on the Model 1000 Control Unit was occasionally triggered inappropriately". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0451-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Kimberly-Clark Corporation
Product
Kimberly-Clark Patient Warming System - Model 1000 Control Unit The product is a thermal regulating system, indicated for monitoring and controlling patient temperature.
Product code
DWJ System, Thermal Regulating
Reason for recall
Error 85 alarm on the Model 1000 Control Unit was occasionally triggered inappropriately.
Root cause tag
design
FDA root cause
Component design/selection
Action
Kimberly Clark sent an Urgent Field Corrective Action letter dated September 16, 2010, to all affected customers. The letter identified the product, the problem, and the aciton to be taken by the customer. Customers were asked to confirm whether their facility as the affected product by faxing the completed Field Corrective Action Response Form to 920-380-6682. Kimberly-Clark will arrange to perform the upgrade at the customer's facility. Customers with questions should call the Field Corrective Action Coordinator at 770-587-7400. For questions regarding this recall call 770-587-8393.
Quantity in commerce
400 units
Distribution
Nationwide Distribution including AZ, AK, CA, CO, DE, DC, FL, GA. ID, IL, IN, IA, KS, KY, LA, MD, ME MA, MS, MO, MT, NE, NJ, NM, NV, NY, NC, OH, OR, PA, RI, SC, TN, TX, VA, WV, and WI
Product codes and lots
Serial Number - SA09043017, SA09043019, SA0904320, SA09043018, SA09043015, SA09043012, SA09043016, SA09043014, SA09043013, SA09043009, SA09043011, SA09043010, SA09043008, SA09043007, SA09043006, SA09043003, SA09043005, SA09043004, SA09043002, SA09043001, SA09020013, SA09020012, SA09020014, SA09020011, SA09020010, SA09020009, SA0902008, SA09020006, SA09020007, SA08343100, SA09020004, SA09020005, SA09020002, SA09020003, SA09020001, SA08343099, SA08343098, SA08343097, SA08343096, SA08343095, SA08343094, SA08343093, SA08343092, SA08343091, SA08343090, SA08343089, SA08343087, SA08343088, SA08343086, SA08343085, SA08343083, SA08343084, SA08343053, SA08343073, SA08288014, SA08288002, SA08343054, SA08343082, SA08343081, SA08343080, SA08343079, SA08343078, SA08343077, SA08165007, SA08343076, SA08343075, SA08343074, SA08343072, SA08343055, SA08288007, SA08343071, SA08343070, SA08343069, SA08343068, SA08343067, SA08343066, SA08343065, SA08343064, SA08343063, SA08343059, SA08343058, SA08343062, SA08343061, SA08343060, SA08343056, SA08343057, SA08288012, SA08288017, SA08288016, SA08288015, SA08288013, SA08288011, SA08288009, SA08288006, SA08288003, SA08288010, SA08288008, SA08288004, SA08288005, SA08288001, SA08165013, SA08165014, SA08165015, SA08165018, SA08165016, SA08165017, SA08165006, SA08165005, SA08165011, SA08165012, SA08165010, SA08165009, SA08165004, SA08165008, SA08165002, SA08165003, SA08144035, SA08144030, SA08165001, SA08144034, SA08144033, SA08144032, SA08144031, SA08144029, SA08144028, SA08144025, SA08144027, SA08144021, SA08144026, SA08144024, SA08144023, SA08144022, SA08136010, SA08136006, SA08136003, SA08136009, SA08136005, SA08136004, SA08136007, SA08136001, SA08136002, SA08136008, SA08094129, SA08094137, SA08094141, SA08094136, SA08094150, SA08094153, SA08094152, SA08094151, SA08094149, SA08094145, SA08094148, SA08094147, SA08094143, SA08094146, SA08094142, SA08094144, SA08094140, SA08094139, SA08094138, SA08094134, SA08094130, SA08094132, SA08094135, SA08094131, SA08094133, SA08094127, SA08094128, SA08094125, SA08094126, SA08094124, SA08017044, SA08017049, SA08017045, SA08017040, SA08017034, SA08017035, SA08017041, SA08017047, SA08017042, SA08017043, SA08017048, SA08017039, SA08017046, SA08017038, SA08017037, SA08017030, SA08017036, SA08017028, SA08017029, SA08017027, SA08017033, SA08017031, SA08017032, SA07344022, SA07344029, SA07344028, SA07344026, SA07344027, SA07344030, SA07344021, SA07344023, SA07344024, SA07344025, SA07320002, SA07320001, SA07320014, SA07320015, SA07320013, SA07320011, SA07320012, SA07320006, SA07320005, SA07320008, SA07320010, SA07320009, SA07320007, SA07320004, SA07320003, SA07261006, SA07257008, SA07261004, SA07261007, SA07261002, SA07257004, SA07257007, SA07257003, SA07257002, SA07261005, SA07261008, SA07261003, SA07261001, SA07257005, SA07257006, SA07257001, SA07250005, SA07250004, SA07250002, SA07250003, SA07250001, SA07228007, SA07228008, SA07228001, SA07228002, SA07228003, SA07228004, SA07228005, SA07228006, SA07180006, SA07180005, SA07180004, SA07180003, SA07180002, SA07180001, SC07171006, SC07171005, SC07171004, SC07171003, SC07171002, SC07171001

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0451-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0451-2013Class IITerminatedVoluntary: Firm initiated